Environnement
S.A
MIR 9000
Duplication prohibited
3–10
JULY 2009
NOTE :
A flowrate of 20 l/h should correspond to 2 or 3 stars The flowrate graph is defined in the
configuration menu at the factory (zero and full scale),
The numerical value of the flowrate is displayed in the «Test / Optical Bench» menu.
3.3.2.1.1 Keys
« Sample »
This switches the analyzer to its SAMPLE gas input. It becomes highlighted after acknowledgement.
« Zero »
This switches the analyzer to its ZERO input. It becomes highlighted after acknowledgement (see also
TIG OPTION).
« Span »
This switches the analyzer to its SPAN input. It becomes highlighted after acknowledgement (see also
TIG OPTION).
« Ref. Z »
This starts a Zero Reference cycle. It becomes highlighted during the first dead time.
« >> » Only if there are more than 3 measurements
This enables selecting the following 3 measurements. The measurements are spaced by +3 each time
the key is pressed, module of the number of measurements programmed.
TIG OPTION
Reassignment of the “Zero” and “Span" keys
If the TIG option is activated in the
«Configuration / Measurement mode»
menu, the ZERO and
SPAN keys change their designation and enable execution of the different commands, specific to the
TIG board.
List of assignments of the “Zero” and “SPAN" keys
”Zero”
“Span”
Zero
Ana
Span1 Probe
Span2
↓
selection of the function associated with the «Zero» key
Span3
Span4
Span5
Summary of Contents for Envea MIR 9000
Page 10: ...Duplication prohibited MIR 9000 EnvironnementS A 0 10 JULY 2009 Page intentionally left blank...
Page 22: ...EnvironnementS A MIR 9000 Duplication prohibited 1 12 JULY 2009 Figure 1 7 Analysis cabinet...
Page 27: ...Duplication prohibited MIR 9000 EnvironnementS A 1 17 JULY 2009 Page intentionally left blank...
Page 35: ...Duplication prohibited MIR 9000 EnvironnementS A 2 5 JULY 2009 Page intentionally left blank...
Page 102: ...EnvironnementS A MIR 9000 Duplication prohibited 4 2 JULY 2009 Page intentionality left blank...
Page 157: ......
Page 159: ......
Page 161: ......
Page 163: ...EnvironnementS A MIR 9000 Duplication prohibited 6 10 JULY 2009 Figure 6 4 Oxygen probe...
Page 165: ...EnvironnementS A MIR 9000 Duplication prohibited 6 12 JULY 2009 Page intentionally left blank...
Page 180: ...EnvironnementS A MIR 9000 Duplication prohibited 6 22 JULY 2009 Page intentionally left blank...
Page 182: ...EnvironnementS A MIR 9000 Duplication prohibited 7 2 JULY 2009 Page left intentionally blank...
Page 206: ...EnvironnementS A MIR 9000 Duplication prohibited 7 28 JULY 2009 Page left intentionally blank...
Page 212: ...EnvironnementS A DAC8 board Duplication prohibited MARCH 2018 6 Page intentionally left blank...