MAINTENANCE SHEET
Serial no. MONITOR:
OPERATION SHEET 4.3.3
Subject: Inspection of the pump
PAGE: 1/2
Frequency: 6 months
Type of opera-
tion
Date
−
Inspection of the diaphragm and the valves
(Figure 4–1)
Maintenance kit: Ref. V02-K-113-909
•
Switch off the analyzer.
•
Disconnect connector J6 from the motherboard.
•
Disconnect the 2 tubes from the yoke (18) of the pump and mark their
respective positions.
•
Loosen the 2 knurled fastening screws and take out the pump by
freeing it from the enclosure located behind it.
•
Mark the relative positions of parts (18), (16), (10) and (5) using a
pencil.
•
Loosen the screw (1).
•
Loosen the 4 screws (19) and remove the head of the pump (make
sure that the articulation washer is not mislaid (20)).
•
Disassemble the various parts, starting with part (16). Check the
condition of the valves (15). Replace them if they are marked too
much.
•
Check the condition of the circlips (14). Replace them if they are
oxidized.
•
Check the condition of the diaphragm (11). Replace it if it is cracked or
warped.
•
Replace the seal (17) after 2 disassembly operations.
•
Clean the internal parts and components of the pump with alcohol.
•
Assemble the unit when it is sufficiently dry. Observe the orientation of
the parts.
•
Replace the suspension rings (4) if necessary.
•
Check the tightening of the screws (3).
NH
3
/ H
2
O version of the MIR :
such a maintenance is pointless
−
Required tools:
•
∅
6 mm flat screwdriver,
•
Alcohol,
•
Paper
tissue,
•
Soft
brush.
−
Type of operation:
R: Replacing
C: Cleaning
I: Inspection
Summary of Contents for Envea MIR 9000
Page 10: ...Duplication prohibited MIR 9000 EnvironnementS A 0 10 JULY 2009 Page intentionally left blank...
Page 22: ...EnvironnementS A MIR 9000 Duplication prohibited 1 12 JULY 2009 Figure 1 7 Analysis cabinet...
Page 27: ...Duplication prohibited MIR 9000 EnvironnementS A 1 17 JULY 2009 Page intentionally left blank...
Page 35: ...Duplication prohibited MIR 9000 EnvironnementS A 2 5 JULY 2009 Page intentionally left blank...
Page 102: ...EnvironnementS A MIR 9000 Duplication prohibited 4 2 JULY 2009 Page intentionality left blank...
Page 157: ......
Page 159: ......
Page 161: ......
Page 163: ...EnvironnementS A MIR 9000 Duplication prohibited 6 10 JULY 2009 Figure 6 4 Oxygen probe...
Page 165: ...EnvironnementS A MIR 9000 Duplication prohibited 6 12 JULY 2009 Page intentionally left blank...
Page 180: ...EnvironnementS A MIR 9000 Duplication prohibited 6 22 JULY 2009 Page intentionally left blank...
Page 182: ...EnvironnementS A MIR 9000 Duplication prohibited 7 2 JULY 2009 Page left intentionally blank...
Page 206: ...EnvironnementS A MIR 9000 Duplication prohibited 7 28 JULY 2009 Page left intentionally blank...
Page 212: ...EnvironnementS A DAC8 board Duplication prohibited MARCH 2018 6 Page intentionally left blank...