STACK 710
4 Getting Started
ENVEA
UK Ltd
(PC-000878-MA Issue 7, December 2021)
39
51
Number of
faults
Indicates the
number of faults
currently present
(regardless of
mask states)
0 - 32
NA
52
Error Flags
Error flags
Bit 0 = Zero motor jammed Bit
1 = Upscale motor jammed Bit
2 = Source LED fail
Bit 3 = Flood Led fail Bit 4 =
ADC over-range Bit 5 = ADC
fault
Bit 12 = Negative opacity Bit
13 = Checksum error
Bit 14 = Lens contamination
limit
Bit 15 = Failure during
calibration
NA
201
Average
during audit
0 = Off
1 = On
NA
202
Initiate
Calibration
Bit 0 = Cal check
Bit 1 = Cal audit
NA
95
Cal check
interval
In hours (Default =
24)
0 - 1000
NA
96
Time to
next cal
Time to next cal in
minutes
0 - 65535
NA
Note: Read operation on an unassigned Register will return zero rather than an exception.