STACK 710
4 Getting Started
ENVEA
UK Ltd
(PC-000878-MA Issue 7, December 2021)
31
Gravimetric
O.D.
Average Optical Density calculated during gravimetric
test
Running In
Positioning the zero reflector or upscale filter to an active position
Running Out
Positioning the zero reflector or upscale filter to an inactive position
4.5
Using the Instrument for the First Time
Tools required: a 19 mm spanner and a 5 mm hexagonal (Allen) key.
Note: When entering numbers using the
and
arrows on the keypad, wait until
the
symbol appears on the display before pressing the
key to confirm
the value.
4.5.1
Complete Installation
Ensure that the installation has been completed and that the purge air blowers
are running.
4.5.2
Check the Pathlength Correction Factor
If the STACK 710 is to be used to report Opacity, the Pathlength Correction Factor
(PLCF) must be set correctly.
PLCF is not relevant for Dust Density or Optical
Density (Extinction) measurements.
Press
Use the
and
arrows to select
Parameters
and press
Use the
and
arrows to select Parameter Number
and press
Use the
and
arrows to change the Parameter number to 1 7 and press
If the PLCF is correct press
three times to return to the main screen.
If it is incorrect, the opacity monitor must be returned to ENVEA for
reprogramming.
Summary of Contents for STACK 710
Page 1: ...STACK 710 USER MANUAL ...
Page 2: ......
Page 3: ...STACK 710 Particulate Measurement System USER MANUAL Publication Part Number PC 000878 MA ...
Page 63: ...STACK 710 ENVEA UK Ltd PC 000878 MA Issue 7 December 2021 63 Notes ...
Page 64: ...STACK 710 64 PC 000878 MA Issue 7 December 2021 ENVEA UK Ltd Notes ...
Page 65: ...STACK 710 ENVEA UK Ltd PC 000878 MA Issue 7 December 2021 65 Notes ...
Page 67: ......