TECHNICAL DATA
Mains voltage
: 115 V
, ± 15% 50-60 Hz
220-240 V
, ± 10% 50-60 Hz
Current consumption
: 0.13 A (230 V
)
Mains fuses
: 2 x T1.6H250V
Patient leakage current
: typically 10 µA (requirement IEC eis < 100 µA)
Ditto, first fault condition
: typically 20 µA (requirement IEC eis < 500 µA)
Earth (ground) leakage current
: typically 30 µA (requirement IEC eis < 500 µA)
Ditto, single fault condition
: typically 60 µA (requirement IEC eis < 1000 µA)
Medical device classification
: IIb (according to Medical Device Directive (93/42/EEC))
Safety class
: I* type BF**, according to IEC 60601-1
Weight
: 5,5 kg
Dimensions
: 36 x 33,5 x 12 cm
Safety tests
: TÜV Rheinland
Environment conditions for transport and storage
Environment temperature
: -10° till +50° C
Relative humidity
: 10 till 95 %
Atmospheric pressure
: 500 till 1060 hPa
Environments conditions for normal use
Environment temperature
: 10° till 40° C
Relative humidity
: 10 till 90 %
Atmospheric pressure
: 500 till 1060 hPa
Classification
Medical class IIb
This equipment complies with all requirements of the Medical Device Directive (93/42/EEC).
International Safety class IEC 60601-1
*I
: The equipment has a safety earth (ground) connection, and must be connected to
an earthed (grounded) wall socket.
**BF : The equipment has a floating patient circuit. The Endomed 581
ID
meets the re-
quirements of IEC 60601-1 and IEC 60601-2-10.
Technical modifications reserved.
ORDERING DATA
For the ordering data of the Endomed 581
ID
, standard accessories and additional accessories we refer to the Cata-
logue for Physiotherapy.
Type BF