Such adjustment should be made without extracting the filler head, to prevent the liquid from
spilling out of the bag.
Using the handwheel as indicate before raise the prong until the liquid inside the tube ebbs back into
the bag. Block the prong in place at this new height and remove the filling head. The level of the
liquid inside the bag must remain within the rigid ring section. The height thus determined will be
valid for all the bags of the same capacity. When the size of the bag varies the height will need to be
readjusted.
4.4 DIFFERENT BAG FORMAT
. To modify the bag format utilised perform the following
corrections:
Check the height of the prong and if necessary adjust in relation to the size of the bag
Set the no. of litres needed to fill the new bag format
Adjust, if necessary, only the COR parameter
Before starting up the machine for the first filling operation press the RESET button for 5
seconds and check that the counter “QUANTITA’ EROGATA” is at 0. If a different value is
shown keep the RESET button pressed until the O figure appears.
All repairs must be carried out directly by Enoitalia srl or by qualified and specialised staff
In the case of any repair operations use original spare parts only by contacting the retailer or
manufacturer directly.
PROBLEM
REMEDY
The filling machine does not stop upon reaching
the no. of litres set.
Check that the compressor is turned on and that
the air tubes are open.
Check that the air tube is connected to the
pneumatic valve.
The filling machine stops after a few litres without
reaching the set level.
Press the reset button and check that all the
counters show the correct start value.
It is difficult to insert the filler head inside the
bags.
Check that the head is centred and adjust using the
screws for the purpose.
The pincer cannot get the cap out.
Check the exact position of the cap and if
necessary adjust it using the handwheel for the
purpose.
Check the load springs of the pincer.
Check that the cap is compatible with the pincer
used and if necessary contact the manufacturer.
The air depression tube fills with liquid.
Raise the fork holding the bags.
If only a matter of a few drops this is not a
problem since they will be expelled from the
tubes at the next filling operation.
Liquid leaks onto the lower shelf.
These are drops of liquid which are expelled from
the air depression outlet. These drops will not
cause any damage to the system, but to collect
them put a recipient under the outlet tube.
Liquid leaks out of the decanting pump.
Check and replace the mechanical pump seal.
WARNING
5. POSSIBLE CAUSES OF MALFUNCTIONING
PULIZIA, MANUTENZIONE RIMESSAGGIO
Summary of Contents for BB 20
Page 17: ......