INSTALLATION INSTRUCTIONS
1.
Connecting the Oxygen Hose
- On the back panel of the unit there is a port marked
"
OXYGEN IN
", connect one end of the
Green Oxygen Hose
to this port and connect the
other end of this hose to your
50 PSI
oxygen source. Since the oxygen fittings are
universally standard, you may use your own oxygen hose if desired.
CAUTION: IT IS EXTREMELY IMPORTANT THAT THE OXYGEN BE REGULATED TO A
PRESSURE OF
50 POUNDS PER SQUARE INCH (PSI)
, FOR THE MINUTE VOLUME PER
KILOGRAM DISPLAY TO READ ACCURATELY. PRESSURE OVER 50 PSI MAY DAMAGE
INTERNAL PARTS VOIDING THE WARRANTY.
2.
Connecting the Vaporizer Hose
- On the back panel of the unit there
is a port marked "
TO VAPORIZER
", connect one end of the non marked silicon hose to this
port and connect the other end to your vaporizer inlet port.
3.
Connecting the Vaporizer Hose
- On the back panel of the unit there is a port marked
"
FROM VAPORIZER
", connect one end of the blue striped silicon hose to this port and
connect the other end to your vaporizer outlet port.
NOTE: The ADS 2000 must be used with a
precision
vaporizer.
4.
Connecting the Scavenger Tubing
- Connect one end of the Blue Scavenger tubing
to the "
SCAVENGER OUT
” port on the back of the unit. Connect the other end to either a
"passive" or "active" scavenging system. This will help to eliminate all traces of anesthetic
gases used in the operating room that could pose a hazard to personnel. The use of a f/air
anesthesia gas filter unit or active anesthesia scavenger system is highly recommended.
Care must be taken to ensure that the scavenger port is not blocked by improper use of
these devices as the animal's ease of expiration depends on the resistance of the
scavenger line.
5.
Connecting the Power Adapter
-
Connect the small male plug of the power
adapter into the back of the ADS 2000 at the 15 VDC @ 1.2 A outlet. Then plug the adapter
into an electrical outlet. The supplied power adapter continuously charges the internal
battery backup.
NOTE: On a full charge the ADS 2000 has 10 to 12 hours of battery backup for
uninterrupted operation during power failure. Complete battery recharging is
accomplished in approximately 8 hours.
6.
Connecting the Breathing Circuit
- Connect the two large ends of the breathing circuit to
the "BREATHING CIRCUIT" ports on the front panel of the ADS 2000. You may use your own
breathing circuit if you desire. The top port (inspiratory) of the ADS 2000 feeds oxygen / anesthesia
to the patient
. The bottom port (expiratory) is the exhaust. After the lungs have been inflated
the internal exhale valve allows the natural elasticity of the animal's lungs to exhale oxygen /
anesthesia out through this port.
6
Engler Engineering Corporation
ADS 2000