Set Mode cont.
c. The default number of
Breaths Per Minute
is 7.5
d. The default
Peak Inspiratory Pressure
is pre-selected to a value of
15 cm. of H
2
O.
e. The
Assist
feature is on and defaults at a pre-selected value of -3.0 cm. of
H
2
O.
Every time you turn the ADS 2000 on, the LCD display should show the
default values
as
shown on page 10, indicating a
successful self-test
.
3. To enter a different weight, simply press either the
WEIGHT UP
or
WEIGHT DOWN
buttons on the front of the ADS 2000 while in
Set Mode
, until the upper line of the LCD
displays the desired weight.
NOTE: Weight values under 10 Kilograms are set to the nearest 0.5 Kilograms, while
weight over 10 Kilograms are set to the nearest 1 Kilogram.
NOTE: It is important to set the unit for the correct weight for each patient so that the
MINUTE VOLUME PER KILOGRAM
value will be calculated correctly.
4. As you select different weight values the ADS 2000 automatically provides default
ventilation parameters. At any time before or during a procedure, you can change any of
following parameters:
a.
LITERS PER MINUTE
- To adjust the
FLOW RATE
to the patient, simply press the
FLOW
RATE UP
or
FLOW RATE DOWN
buttons on the front of the ADS 2000.
b.
BREATHS PER MINUTE
- To adjust the number of
BREATHS PER MINUTE
delivered to the patient, simply press the
BREATHS PER MINUTE UP
or
BREATHS
PER MINUTE DOWN
buttons on the front of the ADS 2000.
c.
PEAK INSPIRATORY PRESSURE
- To adjust the Peak Inspiratory Pressure delivered to
the patient, press either the
P.I.P. UP
or
P.I.P. DOWN
buttons on the front of the ADS
2000.
d.
ASSIST
- To adjust the amount of inspiratory effort needed to initiate a breath by the
patient, press either the
ASSIST UP
or
ASSIST DOWN
buttons on the front of the ADS
2000.
NOTE: To turn the
ASSIST
feature off, press and hold the
ASSIST DOWN
button
until the LCD displays "
OFF
" where the
ASSIST
value was located.
ADS 2000
Engler Engineering Corporation
13