Self-Test cont.
3. If upon
Self-Test
you get the following display:
This indicates that there is leak somewhere in the system.
a. There is debris or condensation trapped inside the unit, perform a
FLUSH
of
the unit by following the instructions in the
UNDERSTANDING THE FLUSH
MODE
section of this manual. Then Press the
WEIGHT DOWN
button to
Retry
.
b.
One of the hoses may be loose, check all connections and Press the WEIGHT
DOWN button to
Retry
.
c. The Breathing Circuit has a leak or cuff is loose, check the Breathing Circuit
for leaks, confirm it is securely connected to the
Breathing Circuit Ports
on the
front of the unit. Press the
WEIGHT DOWN
button to
Retry
.
d. Vaporizer has a leak or hose(s) connected to vaporizer have come loose.
e.
Pop Off Valve
adjustment needed. This must be completed by EEC as
specialized tools and test equipment is required.
Call 800-445-8581
4.
If upon
Self-Test
you get the following display:
This display indicates that there is a minor leak or the
Safety Pop Off
is
releasing at too low a pressure.
ADS 2000
Engler Engineering Corporation
31