Fig.3
6. When the unit passes self test, the LCD Display will show (fig 2 A), its
Automatic Default
–
20 kilograms. The ADS 2000 has just passed the self-test.
7. If you get readout 2 B. “
SWITCH TO SET
", then place the
SET / RUN
switch to "
SET
". The
display will now give readout "A" or "C", if you get readout "A." then proceed to Step 6.
8. If you get readout 2 C. “
SWITCH MASK OFF"
then place the "
MASK
" switch to "O" or
"OFF". The display will now give readout "A", proceed to Step 6.
9.
If you get any one of the error messages "D" through "H" please refer to
TROUBLESHOOTING THE ADS 2000,
section of this manual, beginning on page 24.
10. If a message as in figure 3 is displayed, a hose from the Breathing Circuit or the
Gas
Sampling Circuit
is not connected properly. Re-connect hoses and try again.
The ADS 2000 is ready for operation!
Note: Do not connect TEST LUNG / patient until unit has passed self-test.
SET MODE
In this section you will learn and set up the parameters of the ADS 2000 in the "
SET
MODE
". The "
SET MODE
" is the resting or static mode for the ADS 2000. The "
SET
MODE
" is the mode in which you will enter the patients' weight in kilograms and from
that input the ADS 2000 will select all of the other parameters for you. Of course, the
ADS 2000 will only select values based on an average, if at any time you wish to
change any parameter, you may do so at any time.
THE
LCD DISPLAY
in
SET
MODE
. Figure 2 A shows the default 20 kilograms display
once the self test process is completed.
1. Make sure that the
SET / RUN
switch is in the "
SET MODE
".
2. As shown on figure 2A we can see the following:
a. Patient default
Weight
is pre-selected at 20 Kilograms.
b. The default
Flow Rate
is pre-selected to a value of 24 liters per minute.
12
Engler Engineering Corporation
ADS 2000