USING THE BREATHE FEATURE
1. The
BREATHE
button can be pressed at any time between cycles (breaths) to
initiate a breath, overriding both timed and assisted respiration.
NOTE: To test the
BREATHE
function, turn the ADS 2000
ON
, (see page 10) install
the
TEST LUNG
onto the end of the breathing circuit, place the
SET / RUN
switch to
"
RUN
". When the ADS 2000 is between breaths, press the
BREATHE
button, the
TEST LUNG
will inflate and deflate to the preset parameter.
USING THE MASK MODE
NOTE: Unit must be turned
ON
while in
Normal Mode
(
50 PSI
), then the tubing
configuration changed to
Mask Mode
.
1. The
Mask Adapter
consists of the Y piece and the adapter. Masking a patient
down requires an adapter, which is supplied with the unit.
2. As shown below, connect the Breathing Circuit to the end of the “Y” piece.
3. Attach the “Blue Scavenger Tubing” to the large adapter that is connected to the
“Y” piece.
4. Connect the Gas Sampling Elbow to the large port of the “Y” piece as shown.
Mask Adapter
Gas Sampling
4
Blue Scavenger Tube
3
System
2
To Mask
Breathing Circuit
5. Set the vaporizer to the desired concentration.
6. Place the "
MASK
" switch
ON
or "
I
". The LCD display will look like Figure 6, and a
continuous flow of oxygen with anesthetic gas will flow through the mask at your pre-
selected
FLOW RATE
.
Figure 6
22
Engler Engineering Corporation
ADS 2000