Maintenance
Stamolys CA71CODcr
40
Hauser
7.6
Cleaning
7.6.1
Housing
NOTICE
When cleaning, ensure that you do not damage the nameplate on the analyzer.
►
Do not use any solvent-based cleaning agents.
To clean the analyzer housing, proceed as follows:
• Wipe up any reagent contamination with a disposable cloth and rinse the area with clean water.
Then dry the cleaned areas with a cloth.
• If the hose box has become white, it must be replaced. Cleaning it is no longer an option.
7.6.2
Cleaning of the filter mat of the housing fan
The cover of the filter mat is at the backside of the analyzer (center, top).
1.
Remove the cover to replace the filter mat.
2.
To clean the filter mat you have various options:
a. Wash the filter mat in water (approx. 40
˚
C) with a mild detergent if necessary.
b. Alternatively you can beat or vacuum the mat or blow it with compressed air.
c. In case of greasy substances: Wash the filter mat in benzine, trichlorethylene or warm
water with grease solvent added.
NOTICE
Damage due to residues of organic solvents and incorrectly installed filter mat
Lost of the guaranteed protection degree due to missing, damaged or incorrectly installed filter mat
►
If organic solvents are used, the mat has to be completely dried on air before re-assembled.
►
Always take care to install the filter mat correctly.