Operation
Stamolys CA71CODcr
30
Hauser
5.5
Calibration
Recommendation:
Start offset and standard calibration manually via the CONFIGURATION menu.
First define the
frequency offset and only then define the calibration factor.
You can also view the current calibration data in this menu.
Calibration example 1
Manual calibration
1.
Hold
H
down until AUTO MEASURING appears.
2.
Use
W
to move through the CONFIGURATION menu and press
F
.
3.
Use
V
to enter "3" and press
F
.
4.
Use
F
to go to the "Calibrate Offs?" item and press
J
to confirm.
Once the offset calibration has been performed, the current frequency offset is displayed in
Hertz.
5.
Use
K
to go to the "Calibrate Slope?" item and press
J
to confirm.
Once the standard calibration has been performed, the current calibration factor is displayed.
6.
Press
H
to return to the main menu (AUTO MEASURING).
7.
Press
F
for the measuring mode.
Calibration example 2
Automatic calibration
This consists of two offset measurements and a calibration of the K-factor.
1.
Hold
H
down until AUTO MEASURING appears.
2.
Use
W
to move through the PARAMETER ENTRY menu and press
F
.
3.
Use
F
to go to the "Calib. interval" item.
4.
Use the
W
and
V
keys to set a value > 0.
5.
Press
F
to accept the value and then press
H
twice to return to the main menu (AUTO
MEASURING).
6.
Press the
F
key again. This takes you back to measuring mode.
The calibration is now carried out automatically.