Wiring
Stamolys CA71CODcr
16
Hauser
Version without Modbus
Version with Modbus RS485
Function
Designation
Terminal one channel
Mains
L
3
N
2
PE
1
Alarm value 1, channel 1
COM
25
NC
24
NO
23
Alarm value 2, channel 1
COM
28
NC
27
NO
26
Fault
COM
19
NC
18
NO
17
Reserve (unassigned terminals)
COM
22
NC
21
NO
20
Analog output 1
0/4 ... 20 mA
+
36
–
35
Screen
PE
1
Sample conditioning
remote control
Input
57
0 V
53
Function
Designation
Terminal one channel
Mains
L
3
N
2
PE
1)
1
Alarm value 1
COM
25
NC
24
NO
23
Alarm value 2
COM
28
NC
27
NO
26
Fault
COM
19
NC
18
NO
17
Reserve (unassigned terminals)
COM
22
NC
21
NO
20
Analog output 1
0/4 ... 20 mA
2)
+
I +
–
I -
Shield
*