Stamolys CA71CODcr
Operation
Hauser
25
5.2.4
PARAMETER ENTRY
Menu item
Range of adjustment
(factory settings in bold)
Display
Info
Measuring range
start
COD-A: 0 to 200 mg/l /
0 mg/l
COD-B: 0 to 5000 mg/l /
0 mg/l
The specified concentration is allocated a value of 0 or 4
mA at analog output
1)
.
Measuring range
end
COD-A: 50 to 200 mg/l /
200 mg/l
COD-B: 50 to 5000 mg/l /
5000 mg/l
The specified concentration is allocated a value of 20 mA
at analog output.
Alarm value AV1
COD-A: 0 to 200 mg/l /
50 mg/l
COD-B: 0 to 5000 mg/l /
1000 mg/l
Concentration threshold value for limit relay 1 (switching
hysteresis 2% of alarm value).
Alarm value AV2
COD-A: 0 to 200 mg/l /
150 mg/l
COD-B: 0 to 5000 mg/l /
4000 mg/l
Concentration threshold value for limit relay 2 (switching
hysteresis 2% of alarm value).
Time
1. measurement
01.01.96 00:00 to 31.12.95 23:59
Date format DD.MM.YY, time hh.mm. After each change
the instrument does not wait for the measuring interval. If
the measurement is to start immediately, set the time in
the past.
Measuring interval
10 min
2 min to 1440 min
2 min
Time between two measurements.
If the setting is 2 minutes, the measurements take place
without any pauses.
Time
1. calibration
01.01.96 00:00 to 31.12.95 23:59
Time of 1st calibration (DD.MM.YY, time hh.mm). After
each change the instrument does not wait for the
calibration interval. If the calibration is to start
immediately, set the time in the past.
Analyzers are delivered pre-calibrated.
– Start the 1st calibration 2 hours after the initial startup
at the earliest (warm-up phase)
– Set the time to 8:00 to track the effects of calibration in
the curve.
– If you have started a calibration manually, you should
redefine the time of the 1st calibration because the
interval is calculated starting from the last calibration.
Calibration interval
0 to 720 h
48 h
Time between two calibrations.
The "0 h" setting means that calibration does not take
place.
Recommended: calibration interval of 48 to 72 h.
Calibration solution
COD-A: 1 to 200 mg/l /
100 mg/l
COD-B: 1 to 5000 mg/l /
2500 mg/l
Concentration of the calibration standard.
Select a standard whose concentration is in the top third
of the measuring range.
1)
Option 0 or 4 mA: see "CONFIGURATION".
Range start
0.00 mg/l
Alarm val. 1
50 mg/l
Meas. interval
2 min
Calib. solution
100 mg/l