Stamolys CA71CODcr
Operation
Hauser
27
5.2.8
SERVICE
NOTICE
In this menu, you can trigger direct function controls. If these controls are incorrectly
combined, however, this could damage the analyzer.
►
Using the Service menu requires detailed knowledge of how the analyzer works and operates.
►
Exit the Service menu with the "M" key. All functions will be reset to the delivery state then.
►
When using the manual operation via the Service menu, make sure that chloride containing
sample never gets in contact with reagent R2 in the reactor. Otherwise precipitating silver
chloride can cause film or coating on the optical windows. Measuring errors would be the
consequence.
Option
Display
Info
Pumps and valves
"Virtual switching board"
Various valve and pump combinations can be selected.
The setting options are:
– Valve 1 and Valve 5:
P (sample/rinse water outlet) or S (standard/chromate
waste outlet)
– Valve 2:
S (standard) or 0 (standard 0 mg/l O2)
– Valve 3 :
P (sample) or H (acid)
– Valve 4 (to empty the reactor):
c (closed) or o (open)
– Pump 1 (sample or standard/ (standard 0 mg/l O2))
and pump 2 (reagents):
s (stop) or g (go)
The following valve combinations are possible:
– V1: P, V2: S, V3:P
Passage for the sample. This combination is
automatically reset on leaving the service menu.
– V1: S, V2: S, V3:P
Passage for standard solution
– V1: S, V2: 0, V3:P
Passage for standard 0 mg/l O
2
– V1: P, V2: S, V3:H
Passage for reagent 1
Heating and cooling
He = Heating
Fa = Cooling ventilator on reactor
Co = Peltier Cooling
Air = Compressor (air supply to reactor)
Signal frequency
Signal frequency of the photometer
6000 Hz