9.2
Simulation
Simulation can be used to check the FMC 671 Z/676 Z and external downstream devices.
Four modes are possible:
•
Simulation of frequency, V9H6
•
Simulation of level, V9H7
•
Simulation of volume, V9H8
•
Simulation of current, V9H9.
When a value is entered at the appropriate matrix, the analog outputs are fed with the
appropriate current and voltage.
Activation
Enter operating mode 6 for channel 1 or 7 for channel 2 in V8H0; the red alarm LED
flashes during the simulation.
Step
Matrix
Entry
Significance
1
V8H0
e.g. 6
Operating mode 6, simulation channel 1
2
-
»E«
Registers entry
Frequency simulation
Step
Matrix
Entry
Significance
1
V9H6
e.g. 100
Depending upon the calibration and linearization, a value
corresponding to 100 Hz is displayed
2
-
»E«
Registers entry
Level simulation
Step
Matrix
Entry
Significance
1
V9H7
e.g. 10
Depending on the calibration and linearization, the analogue
outputs accept values according to the unit in V0V0
2
-
»E«
Registers entry
Volume simulation
Step
Matrix
Entry
Significance
1
V9H8
e.g. 100
Depending on the calibration and linearization, the analogue
outputs accept values according to the unit in V0H0
2
-
»E«
Registers entry
Current simulation
Step
Matrix
Entry
Significance
1
V9H9
e.g. 16
The analogue output is fed with a current of 16 mA and
the corresponding measured value displayed
2
-
»E«
Registers entry
Deactivation
Step
Matrix
Entry
Significance
1
V8H0
e.g. 0
Operating mode 0, standard measuring
2
-
»E«
Registers entry
Silometer FMC 671 Z/676 Z
Chapter 9: Trouble-Shooting
53