RID14
Technical Data
Hauser
39
13.4.5 Humidity
• Condensation permitted as per IEC 60 068-2-33
• Max. rel. humidity: 95% as per IEC 60068-2-30
13.4.6 Degree of protection
IP67. NEMA 4X.
13.4.7 Shock and vibration resistance
10 to 2
000 Hz for 5g as per IEC 60 068-2-6
13.4.8 Electromagnetic compatibility (EMC)
CE EMC conformity
The device complies with all the requirements of IEC 61326-1:2006 and NAMUR
NE21:2007.
This recommendation is a consistent determination as to whether the devices used in
laboratories and in distributed control systems are immune to interference, thus increasing
their functional safety.
ESD (electrostatic discharge)
IEC 61000-4-2
6 kV cont., 8 kV air
Electromagnetic fields
IEC 61000-4-3
0.08 to 4 GHz
10 V/m
Burst (fast transients)
IEC 61000-4-4
1 kV
Surge
IEC 61000-4-5
1 kV asym.
Conducted RF
IEC 61000-4-6
0.01 to 80 MHz
10 V
13.4.9 Measuring category
Measuring category II as per IEC 61010-1. The measuring category is provided for
measuring on power circuits that are directly connected electrically with the low-voltage
network.
13.4.10 Degree of contamination
Pollution degree 2 as per IEC 61010-1.