Technical Data
RID14
38
Hauser
13.2.2 Supply voltage
Voltage is supplied via the fieldbus.
U = 9 to 32 V
DC
, polarity-independent (max. voltage U
b
= 35 V).
13.2.3 Mains voltage filter
50/60 Hz
13.2.4 Current consumption
≤ 11 mA
13.2.5 Cable entry
The following cable entries are available:
• Thread NPT1/2
• Thread M20
• Thread G1/2
13.3 Installation
13.3.1 Orientation
No restrictions, the orientation depends on the readability of the display.
13.3.2 Mounting location
Wall or pipe mounting (see "Accessories" )
13.4 Environment
13.4.1 Ambient temperature limits
–40 to +80 °C (–40 to +176 °F)
The display can react slowly at temperatures < –20 °C (–4 °F).
The readability of the display is no longer guaranteed at temperatures <
–30 °C (–22 °F).
13.4.2 Storage temperature
–40 to +80 °C (–40 to +176 °F)
13.4.3 Altitude
Up to 4
000 m (13
100 ft) above mean sea level in accordance with IEC 61010-1, CSA
1010.1-92
13.4.4 Climate class
According to IEC 60654-1, Class C