Wiring
RIA15
14
Hauser
Terminal
Description
Auxiliary terminals (electrically connected internally)
Potential equalization
• Panel-mounted device:
Terminal on the rear of the housing
• Field device:
Terminal in the housing
Connection without backlighting
Connection with backlighting
Connection with transmitter power
supply and transmitter
LED
-
+
-
-
+
+
I
Y
1
A0017704
1 Transmitter power supply
LED
-
+
+
+
I
Y
1
-
-
A0017705
1 Transmitter power supply
Connection with transmitter power
supply and transmitter using the auxiliary
terminal
LED
-
+
-
-
+
+
I
Y
1
A0017706
1 Transmitter power supply
LED
-
+
+
+
I
Y
1
-
-
A0017707
1 Transmitter power supply
Connection without transmitter power
supply directly in the 4 to 20 mA circuit
LED
-
+
+
2
-
I
Y
A0017708
2 4 to 20 mA power source
LED
-
+
+
2
-
I
Y
A0017709
2 4 to 20 mA power source
NOTICE
SELV Class 2 device
►
The device may only be powered by a power unit with an energy-limited circuit in
accordance with IEC 61010-1: 'SELV or Class 2 circuit'.