Operation
Prosonic S FMU95
32
E Hauser
LEDs
Keys (softkey operation)
The function of the keys depends on the current position within the operating menu (softkey
functionality). The key functions are indicated by softkey symbols in the bottom line of the display.
LED indicating the operating state (pos. 3 in the figure)
green
normal measuring mode; no error detected
red (flashing)
Warning:
An error is detected but the measurement continues. Reliability of the measured value is no longer
ensured.
red
Alarm:
An error is detected. The measurement is interrupted. The measured value assumes the value
specified by the user (parameter "output on alarm").
off
supply voltage missing
Symbol
Meaning
Move downwards
Moves the marking bar downwards within a selection list.
Move upwards
Moves the marking bar upwards within a selection list.
Enter
• Opens the marked submenu, the marked parameter set or the marked parameter
• Confirms the edited parameter value
Previous parameter set
Reopens the previous parameter set within the submenu.
Next parameter set
Opens the next parameter set within the submenu.
Confirm selection
Selects the option of a selection list which is currently marked by the bar.
Increase value
Increases the active digit of an alphanumeric parameter.
Decrease value
Decreases the active digit of an alphanumeric parameter
Error list
Opens the list of all errors which are currently detected.
If a warning is present, this symbol flashes.
If an alarm is present, the symbol is displayed continuously.
Change Display
Change to the next page of measured values (only available if more than one pages of measured
values have been defined; see "display" menu)
Info
Opens the Shortcut Menu, which contains the most important information about the current state
of the instrument
Menu
Opens the Main Menu, which contains
all