Technical data
Proline t-mass A 150 HART
98
Hauser
Current output
Failure mode
Can be selected (as per NAMUR recommendation NE 43)
Minimum alarm
3.6 mA
Maximum alarm
22 mA
Adjustable value
3.6 to 22.5 mA
Pulse/frequency/switch output
Pulse output
Failure mode
Choose from:
• Actual value
• No pulses
Frequency output
Failure mode
Choose from:
• Actual value
• Defined value: 0 to 1250 Hz
• 0 Hz
Switching output
Failure mode
Choose from:
• Current status
• Open
• Closed
Local display
Plain text display
With information on cause and remedial measures
Status signal as per NAMUR recommendation NE 107
Operating tool
• Via digital communication: HART protocol
• Via service interface
Plain text display
With information on cause and remedial measures
Low flow cut off
The switch point for low flow cut off is programmable.
Galvanic isolation
The following connections are galvanically isolated from each other:
• Outputs
• Power supply
Protocol-specific data
HART
Manufacturer ID
0x11
Device type ID
0x66
HART protocol revision
6.0
Device description files (DTM,
DD)
Information and files under: