Proline t-mass A 150 HART
Diagnostics and troubleshooting
Hauser
87
410
Data transfer
1. Check connection.
2. Retry data transfer.
F
Alarm
*
411
Upload/download
1. Check connection.
2. Retry data transfer.
F
Alarm
*
411
Up-/download active
Up-/download active, please wait
C
Warning
*
431
Trim
Carry out trim.
C
Warning
*
437
Incompatible
configuration
1. Restart device.
2. Contact service.
F
Alarm
*
437
Incompatible
configuration
1. Transfer data or reset device.
2. Contact service.
C
Alarm
438
Dataset
1. Check data set file.
2. Check device configuration.
3. Up- and download new
configuration.
M
Warning
*
441
Current output
1. Check process.
2. Check current output settings.
S
Warning
*
442
Frequency output
1. Check process.
2. Check frequency output setting.
S
Warning
*
443
Pulse output
1. Check process.
2. Check pulse output setting.
S
Warning
*
453
Flow override
Deactivate flow override.
C
Warning
*
484
Simulation failsafe
mode
Deactivate simulation.
C
Alarm
485
Simulation process
variable
Deactivate simulation.
C
Warning
*
491
Simulation current
output
Deactivate simulation.
C
Warning
*
492
Frequency simulation
Deactivate simulation.
C
Warning
*
493
Simulation pulse
output
Deactivate simulation.
C
Warning
494
Switch output
simulation
Deactivate simulation.
C
Warning
*
Diagnostic behavior can be changed: Section 12.4 "Adapting the diagnostic behavior"
Diagnosti
c number
Short text
Remedy information
Status
signal
from the
factory
Diagnostic behavior
from the factory
Diagnostics for the process
832
Ambient temperature
Reduce ambient temperature.
S
Warning
*
833
Ambient temperature
Increase ambient temperature.
S
Warning
*
834
Process temperature
Reduce process temperature.
S
Warning
*
835
Process temperature
Increase process temperature.
S
Warning
*
Summary of Contents for Proline t-mass A 150
Page 133: ......
Page 134: ...www addresses endress com...