The assignment of the measured variables to the dynamic variables can be modified and
assigned as desired via local operation and the operating tool using the following
parameters:
• Expert → Communication → HART output → Output → Assign PV
• Expert → Communication → HART output → Output → Assign SV
• Expert → Communication → HART output → Output → Assign TV
• Expert → Communication → HART output → Output → Assign QV
The following measured variables can be assigned to the dynamic variables:
Measured variables for PV (primary dynamic variable)
• Mass flow
• Corrected volume flow
• FAD volume flow
• Temperature
Measured variables for SV, TV, QV (secondary, tertiary and quaternary dynamic
variable)
• None
• Mass flow
• Corrected volume flow
• FAD volume flow
• Temperature
• Totalizer
9.3
Other settings
In the
Configuration
submenu, you can configure other settings for the HART protocol
(e.g. Burst mode)
Navigation path
"Expert" menu → Communication → HART output → Configuration