Description of device functions
Proline Prosonic Flow 91
98
Hauser
10.13
Group LINER
Functional description LINER
LINER MATERIAL
Use this function to display the liner material of the pipe.
The liner material must be specified if a liner is present.
Options
:
LINER NONE
MORTAR
RUBBER
TAR EPOXY
OTHER
Factory setting
:
LINER NONE
SOUND VELOCITY
LINER
Use this function to display the sound velocity of the liner. This is specified by the option
selected in the function LINER MATERIAL. If you edit the predetermined value the liner
material will be reset to the option OTHERS.
The sound velocity of the liner must be entered if the option OTHER was selected in the
function LINER MATERIAL.
User input
:
Fixed-point number 800 to 6500 m/s
Factory setting
:
Depends on the setting selected in the function LINER MATERIAL
LINER THICKNESS
Use this function to enter the thickness of the liner.
User input
:
Fixed-point number 0.0 to 99.9 mm
Factory setting
:
0 mm