Easy-to-clean flowmeter with self-drainable single-tube system and an ultra-

compact transmitter

Application

• Measuring principle operates independently of physical fluid

properties such as viscosity or density

• Dedicated for applications requiring optimal cleanability

under hygienic conditions

Device properties

• Large range of hygienic process connections

• 3A and EHEDG conform

• Fast recovery from CIP/SIP

• Robust, ultra-compact transmitter housing

• High ingress protection: IP69K

• Local display available

Your benefits

• Low installation costs – compact, horizontal installation

thanks to fully self-draining tube

• Fewer process measuring points – multivariable

measurement (flow, density, temperature)

• Space-saving installation – no in/outlet run needs

• Space-saving transmitter – full functionality on smallest

footprint

• Time-saving local operation without additional software and

hardware – integrated web server

• Integrated verification – Heartbeat Technology™

Products

Solutions

Services

Technical Information

Proline Promass S 100

Coriolis flowmeter

TI01037D/06/EN/04.14

71254125

Summary of Contents for Proline Promass S 100

Page 1: ...smitter housing High ingress protection IP69K Local display available Your benefits Low installation costs compact horizontal installation thanks to fully self draining tube Fewer process measuring po...

Page 2: ...nting Safety Barrier Promass 100 40 Environment 40 Ambient temperature range 40 Storage temperature 42 Climate class 42 Degree of protection 42 Shock resistance 42 Vibration resistance 42 Interior cle...

Page 3: ...Proline Promass S 100 Endress Hauser 3 Registered trademarks 76...

Page 4: ...hing any other connections Equipotential connection A connection that has to be connected to the plant grounding system This may be a potential equalization line or a star grounding system depending o...

Page 5: ...ow causes deceleration of the oscillation at the inlet of the tubes 2 and acceleration at the outlet 3 1 2 3 A0016772 The phase difference A B increases with increasing mass flow Electrodynamic sensor...

Page 6: ...rosoft Internet Explorer Also for device version with EtherNet IP output Via Web browser e g Microsoft Internet Explorer Via Add on Profile Level 3 for automation system from Rockwell Automation Via E...

Page 7: ...nd Zone 1 Div 1 Safety IT security We only provide a warranty if the device is installed and used as described in the Operating Instructions The device is equipped with security mechanisms to protect...

Page 8: ...max G max F max G can never be greater than max F G Gas density in kg m at operating conditions DN x mm in kg m3 8 60 15 80 25 1 90 40 1 90 50 2 90 To calculate the measuring range use the Applicator...

Page 9: ...tput Version Passive open collector Maximum input values DC 30 V 25 mA Voltage drop For 25 mA DC 2 V Pulse output Pulse width Adjustable 0 05 to 2000 ms Maximum pulse rate 10000 Impulse s Pulse value...

Page 10: ...EIA TIA 485 A standard Terminating resistor For device version used in non hazardous areas or Zone 2 Div 2 integrated and can be activated via DIP switches on the transmitter electronics module For d...

Page 11: ...odbus RS485 Failure mode Choose from NaN value instead of current value Last valid value EtherNet IP Device diagnostics Device condition can be read out in Input Assembly Local display Plain text disp...

Page 12: ...A 72 B Uo 16 24 V Io 623 mA Po 2 45 W For IIC Lo 92 8 H Co 0 433 F Lo Ro 14 6 H For IIB Lo 372 H Co 2 57 F Lo Ro 58 3 H The gas group depends on the sensor and nominal diameter For an overview and for...

Page 13: ...c variable Mass flow Volume flow Corrected volume flow Density Reference density Temperature Totalizer 1 Totalizer 2 Totalizer 3 The range of options increases if the measuring device has one or more...

Page 14: ...ectronic temperature Oscillation frequency Oscillation amplitude Frequency fluctuation Oscillation damping Tube damping fluctuation Signal asymmetry Exciter current Digital input 1 to 2 Partially fill...

Page 15: ...s 16 Write multiple registers 23 Read write multiple registers Broadcast messages Supported by the following function codes 06 Write single registers 16 Write multiple registers 23 Read write multiple...

Page 16: ...evel Ring DLR No Fix Input RPI 5 ms to 10 s factory setting 20 ms Exclusive Owner Multicast Instance Size byte Instance configuration 0x68 398 O T configuration 0x66 64 T O configuration 0x64 44 Exclu...

Page 17: ...asuring device has one or more application packages Fix Output Output Assembly Activation of reset totalizers 1 3 Activation of pressure compensation Activation of reference density compensation Activ...

Page 18: ...ug 1 Connection version 4 20 mA HART pulse frequency switch output 1 1 Signal transmission pulse frequency switch output 1 2 Signal transmission 4 20 mA HART 1 3 Supply voltage 2 Connection version Mo...

Page 19: ...G Option U plug M12x1 thread M20 Options A B C Device plug Device plug Option Q 2 x plug M12x1 Order code for Housing Option A compact coated aluminum Option B compact hygienic stainless Option C ult...

Page 20: ...ad NPT Options A B Device plug Terminals Option L plug M12x1 thread NPT Option N plug M12x1 coupling M20 Option P plug M12x1 thread G Option U plug M12x1 thread M20 Options A B C Device plug Device pl...

Page 21: ...12x1 thread NPT Option N plug M12x1 coupling M20 Option P plug M12x1 thread G Option U plug M12x1 thread M20 Options A B C Device plug Device plug Option Q 2 x plug M12x1 Order code for Housing Option...

Page 22: ...1 Option C thread G Option D thread NPT A B C Device plug Option I plug M12x1 Order code for Housing Option A compact coated aluminum Option B compact hygienic stainless Option C ultra compact hygieni...

Page 23: ...M12x1 thread NPT Option N plug M12x1 coupling M20 Option P plug M12x1 thread G Option U plug M12x1 thread M20 Options A B C Device plug Device plug Option Q 2 x plug M12x1 Order code for Housing Optio...

Page 24: ...4 20 mA HART pulse frequency switch output 18 PROFIBUS DP 20 Modbus RS485 21 EtherNet IP 23 Supply voltage For all connection versions except MODBUS RS485 intrinsically safe device side Device plug MO...

Page 25: ...ce plug for signal transmission device side 3 2 4 1 5 A0016811 Pin Assignment Coding Plug socket 1 B Socket 2 A PROFIBUS DP 3 4 B PROFIBUS DP 5 Grounding shielding Recommended plug Binder series 763 p...

Page 26: ...99 3729 810 04 Phoenix part no 1543223 SACC M12MSD 4Q When using the device in a hazardous location Use a suitably certified plug Supply voltage Transmitter For device version with all communication...

Page 27: ...M Modbus RS485 for use in intrinsically safe areas 230 mA 10 A 0 8 ms Power supply failure Totalizers stop at the last value measured Depending on the device version the configuration is retained in t...

Page 28: ...67 4 Resistor for HART communication 250 observe maximum load 9 5 Analog display unit observe maximum load 9 6 Transmitter Pulse frequency output 1 _ 12345 2 3 A0016801 9 Connection example for pulse...

Page 29: ...specifications 32 3 Distribution box 4 Transmitter If baud rates 1 5 MBaud an EMC cable entry must be used and the cable shield must continue as far as the terminal wherever possible Modbus RS485 2 1...

Page 30: ...able shield observe cable specifications 32 3 Safety Barrier Promass 100 4 Observe cable specifications 32 5 Non hazardous area 6 Non hazardous area and Zone 2 Div 2 7 Intrinsically safe area 8 Transm...

Page 31: ...current input e g PLC Prerequisite automation system with HART version 6 HART commands 113 and 114 can be processed 2 Cable shield observe cable specifications 32 3 Resistor for HART communication 25...

Page 32: ...34 mm2 22 AWG Cable type Twisted pairs Loop resistance 110 km Signal damping Max 9 dB over the entire length of the cable cross section Shielding Copper braided shielding or braided shielding with fo...

Page 33: ...m cable length mm2 AWG m ft 0 5 20 70 230 0 75 18 100 328 1 0 17 100 328 1 5 16 200 656 2 5 14 300 984 Performance characteristics Reference operating conditions Error limits based on ISO 11631 Water...

Page 34: ...0 700 140 US units DN 1 1 1 10 1 20 1 50 1 100 1 500 inch lb min lb min lb min lb min lb min lb min 73 50 7 350 3 675 1 470 0 735 0 147 238 9 23 89 11 95 4 778 2 389 0 478 1 661 5 66 15 33 08 13 23 6...

Page 35: ...e is a difference between the temperature for zero point adjustment and the process temperature the typical measured error of the sensor is 0 0002 of the full scale value C 0 0001 of the full scale va...

Page 36: ...Value measured value ZeroPoint zero point stability Calculation of the maximum measured error as a function of the flow rate Flow rate Maximum measured error in o r ZeroPoint BaseAccu 100 A0021332 Bas...

Page 37: ...ipe Highest point of a pipeline Directly upstream of a free pipe outlet in a down pipe A0023344 Installation in down pipes However the following installation suggestion allows for installation in an o...

Page 38: ...may reduce the ambient temperature To maintain the minimum ambient temperature for the transmitter this orientation is recommended 2 Applications with high process temperatures may increase the ambien...

Page 39: ...mm 298 402 542 750 1019 B mm 33 33 33 36 5 44 1 C mm 28 28 38 56 75 US units DN in 1 1 2 A in 11 73 15 83 21 34 29 53 40 12 B in 1 3 1 3 1 3 1 44 1 74 C in 1 1 1 1 1 5 2 2 2 95 Zero point adjustment A...

Page 40: ...fety Barrier Promass 100 40 to 60 C 40 to 140 F If operating outdoors Avoid direct sunlight particularly in warm climatic regions Weather protection covers can be ordered from Endress Hauser see Acces...

Page 41: ...a maximum medium temperature Tm 392 F Tm 392 F Ex nA CCSAUS NI SI units Order code for Housing Ta C T6 85 C T5 100 C T4 135 C T3 200 C T2 300 C T1 450 C Option A Compact coated alu Option B Compact hy...

Page 42: ...is immediately greater than or equal to the measured maximum ambient temperature Tma that is present Ta 50 C The row showing the maximum medium temperature is determined 3 Select the maximum medium te...

Page 43: ...cable shield must continue as far as the terminal wherever possible For details refer to the Declaration of Conformity Process Medium temperature range Sensor 50 to 150 C 58 to 302 F Seals No internal...

Page 44: ...50 0 50 C 200 150 100 50 0 50 400 0 10 20 30 40 bar psi 200 400 600 0 A0020971 EN 23 With flange material 1 4404 316 316L Process connection to DIN 11851 300 F 350 250 200 150 100 50 0 50 C 200 150 10...

Page 45: ...DIN 11864 2 Form A DN 8 40 DN 50 300 F 350 250 200 150 100 50 0 50 C 200 150 100 50 0 50 400 0 10 20 30 40 bar psi 200 400 600 0 A0020968 EN 26 With flange material 1 4404 316L Process connection to...

Page 46: ...cess characteristics e g with corrosive process fluids we recommend the use of sensors whose secondary containment is equipped with special pressure monitoring connections order code for Purge connect...

Page 47: ...liquids do not outgas This is prevented by means of a sufficiently high system pressure For this reason the following mounting locations are recommended At the lowest point in a vertical pipe Downstre...

Page 48: ...re of Ta 60 C 140 F NOTICE The insulation can also be thicker than the recommended insulation thickness Prerequisite The temperature at the lower end of the transmitter housing does not exceed 80 C 17...

Page 49: ...191 312 17 60 121 92 3 40 136 147 5 93 5 54 215 388 26 00 173 132 3 50 136 147 5 93 5 54 226 467 40 50 241 167 3 1 If using an extension neck for the extended temperature range order code for Sensor...

Page 50: ...Dimensions SI units DN mm A mm B mm C mm D mm E 2 mm F 1 2 mm G mm H mm K mm L mm 8 133 5 136 8 78 58 8 186 294 8 31 108 92 3 15 133 5 136 8 78 58 8 186 294 12 00 108 92 3 25 133 5 136 8 78 58 8 186...

Page 51: ...connection Order code for Housing option C Ultra compact hygienic stainless M12 device plug A D C B E H F G K L A0019438 Dimensions SI units DN mm A mm B mm C mm D mm E 2 mm F 1 2 mm G mm H mm K mm L...

Page 52: ...in 2 If using a display order code for Display Operation option B values 0 55 in 3 dependent on respective process connection Process connections in SI units Flange connections EN DIN C D A B E L 1 5...

Page 53: ...s connection option AAW Surface roughness flange Ra 3 2 to 6 3 m DN mm A mm B mm C mm D mm E mm L mm 8 1 90 60 3 4 15 7 17 1 15 7 336 15 90 60 3 4 15 7 17 1 15 7 440 25 110 79 4 4 15 7 17 6 26 7 580 4...

Page 54: ...nection option NEW Surface roughness flange Ra 3 2 to 6 3 m DN mm A mm B mm C mm D mm E mm L mm 8 1 95 70 4 15 16 0 15 336 15 95 70 4 15 16 0 15 440 25 125 90 4 19 17 5 25 580 40 140 105 4 19 20 0 40...

Page 55: ...l option LP 1 Tri Clamp 1 4435 316L order code for Process connection option FNW DN mm Clamp in A mm B mm L mm 8 1 50 4 22 10 362 15 1 50 4 22 10 466 3A version available order code for Additional app...

Page 56: ...L mm 8 Rd 34 16 362 15 Rd 34 16 466 25 Rd 52 26 606 40 Rd 65 38 825 50 Rd 78 50 1107 3A version available order code for Additional approval option LP Rd 28 1 8 Threaded hygienic connection DIN 11851...

Page 57: ...IN 11864 2 Form A flange with groove C D A B E X X L 1 5 0 06 2 0 0 08 A0015627 37 Detail X Asymmetrical process connection the part shown in gray is provided by the supplier Engineering unit mm in DI...

Page 58: ...m A 1 4435 316L order code for Process connection option KMW DN mm A mm B mm L mm 8 34 0 16 05 370 15 34 0 16 05 474 25 50 5 26 05 614 40 64 0 38 05 825 50 77 5 50 05 1096 3A version available order c...

Page 59: ...oval option LP ISO 2852 clamp connection L 1 5 0 06 2 0 0 08 A B A0015625 40 Engineering unit mm in Clamp connection ISO 2852 1 4435 316L order code for Process connection option JSB DN mm A mm B mm L...

Page 60: ...1107 3A version available order code for Additional approval option LP SMS 1145 threaded hygienic connection L 1 5 0 06 2 0 0 08 A B A0015628 42 Engineering unit mm in Threaded hygienic connection SM...

Page 61: ...37 4 0 62 0 67 0 62 13 23 3 54 2 37 4 0 62 0 67 0 62 17 32 1 4 33 3 13 4 0 62 0 69 1 05 22 83 1 4 92 3 87 4 0 62 0 73 1 61 31 26 2 5 91 4 75 4 0 75 0 99 2 07 42 17 1 DN with DN flanges as standard Fla...

Page 62: ...l option LP 1 Tri Clamp 1 4435 316L order code for Process connection option FNW DN in Clamp in A in B in L in 1 1 98 0 87 14 25 1 1 98 0 87 18 35 3A version available order code for Additional approv...

Page 63: ...16L order code for Process connection option SAW DN in A in B in L in Rd 40 0 89 14 25 Rd 40 0 89 18 35 1 Rd 40 0 89 23 86 1 Rd 60 1 4 32 64 2 Rd 70 1 91 43 58 3A version available order code for Addi...

Page 64: ...3 23 204 8 04 25 1 NPT 25 0 98 82 3 23 348 13 54 40 1 NPT 45 1 77 102 4 02 526 20 70 50 2 NPT 58 2 28 119 5 4 70 763 30 04 Weight Compact version Weight in SI units All values weight refer to devices...

Page 65: ...1 5 3 Adapter for cable entry with internal thread G or NPT Order code for Housing option A Compact coated aluminum The various cable entries are suitable for hazardous and non hazardous areas Cable e...

Page 66: ...thout internal seals Safety Barrier Promass 100 Housing Polyamide Process connections Flanges EN 1092 1 DIN 2501 EN 1092 1 DIN 2512N ASME B16 5 JIS B2220 Tri Clamp OD tubes Clamp with aseptic connecti...

Page 67: ...Local display The local display is only available with the following device version Order code for Display Operation option B 4 line via communication Display element 4 line liquid crystal display wi...

Page 68: ...tool with COM DTM CDI Communication TCP IP 5 Ethernet switch Via PROFIBUS DP network This communication interface is present in the following device version Order code for Output option L PROFIBUS DP...

Page 69: ...Web server or with FieldCare operating tool with COM DTM CDI Communication TCP IP 3 Standard Ethernet connecting cable with RJ45 plug PROFIBUS DP 1 2 3 A0021270 49 Connection for order code for Output...

Page 70: ...ion interface is present in the following device version Order code for Output option M Modbus RS485 1 2 3 A0016925 1 Service interface CDI of the measuring device 2 Commubox FXA291 3 Computer with Fi...

Page 71: ...ng device is certified and registered by the PROFIBUS User Organization PNO The measuring system meets all the requirements of the following specifications Certified in accordance with PROFIBUS PA Pro...

Page 72: ...EMC of industrial process and laboratory control equipment NAMUR NE 32 Data retention in the event of a power failure in field and control instruments with microprocessors NAMUR NE 43 Standardization...

Page 73: ...ntervals in accordance with operator s risk assessment Concentration Package Description Concentration measurement and special density Calculation and outputting of fluid concentrations Many applicati...

Page 74: ...r HART and FOUNDATION Fieldbus devices in the non Ex area For details see Operating Instructions BA01202S Field Xpert SFX370 Field Xpert SFX370 is a mobile computer for commissioning and maintenance I...

Page 75: ...erating Instructions BA00247R iTEMP The temperature transmitters can be used in all applications and are suitable for the measurement of gases steam and liquids They can be used to read in the fluid t...

Page 76: ...e part sets Specified for each individual accessory 73 Registered trademarks HART Registered trademark of the HART Communication Foundation Austin USA PROFIBUS Registered trademark of the PROFIBUS Use...

Page 77: ......

Page 78: ......

Page 79: ......

Page 80: ...www addresses endress com...

Reviews: