Proline Promass I 100
Hauser
9
Maximum possible full scale value:
max(G)
=
max(F)
· ρ
G
: x = 70
000 kg/h · 60.3 kg/m³ : 90 kg/m³ = 46
900 kg/h
Recommended measuring range
Operable flow range
Over 1000 : 1.
Flow rates above the preset full scale value are not overridden by the electronics unit, with the result
that the totalizer values are registered correctly.
Output
Output signal
Current output
Current output
4-20 mA HART (active)
Maximum output values
• DC 24 V (no flow)
• 22.5 mA
Load
0 to 700 Ω
Resolution
0.38 µA
Damping
Adjustable: 0.07 to 999 s
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
The range of options increases if the measuring device has one or more
application packages.
Pulse/frequency/switch output
Function
Can be set to pulse, frequency or switch output
Version
Passive, open collector
Maximum input values
• DC 30 V
• 25 mA
Voltage drop
For 25 mA: ≤ DC 2 V
Pulse output
Pulse width
Adjustable: 0.05 to 2
000 ms
Maximum pulse rate
10
000 Impulse/s
Pulse value
Adjustable
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
Frequency output
Output frequency
Adjustable: 0 to 10
000 Hz
Damping
Adjustable: 0 to 999 s
Pulse/pause ratio
1:1