Proline Promass I 100
20
Hauser
Depending on the housing version, the transmitters can be ordered with terminals or device plugs.
Order code for
"Housing"
Connection methods available
Possible options for order code
"Electrical connection"
Outputs
Power
supply
Options
A
,
B
Terminals
Terminals
• Option
A
: coupling M20x1
• Option
B
: thread M20x1
• Option
C
: thread G ½"
• Option
D
: thread NPT ½"
Options
A
,
B
Device plug
Terminals
• Option
L
: plug M12x1 + thread NPT ½"
• Option
N
: plug M12x1 + coupling M20
• Option
P
: plug M12x1 + thread G ½"
• Option
U
: plug M12x1 + thread M20
Options
A
,
B
,
C
Device plug
Device plug
Option
Q
: 2 x plug M12x1
Order code for "Housing":
• Option
A
: compact, coated aluminum
• Option
B
: compact, hygienic, stainless
• Option
C
: ultra compact, hygienic, stainless, M12 device plug
L
L
26
27
+
_
24
25
1
2
+
_
+
_
1
2
3
A0016888
2
Terminal assignment 4-20 mA HART with pulse/frequency/switch output
1
Power supply: DC 24 V
2
Output 1: 4-20 mA HART (active)
3
Output 2: pulse/frequency/switch output (passive)
Order code for
"Output"
Terminal number
Power supply
Output 1
Output 2
2 (L-)
1 (L+)
27 (–)
26 (+)
25 (–)
24 (+)
Option
B
DC 24 V
4-20 mA HART (active)
Pulse/frequency/switch
output (passive)
Order code for "Output":
Option
B
: 4-20 mA HART with pulse/frequency/switch output