Proline Promass I 100
Hauser
13
Transmitter
Intrinsically safe values
Order code for
"Approvals"
Terminal numbers
Supply voltage
Signal transmission
20 (L-)
10 (L+)
62 (A)
72 (B)
• Option
BM
: ATEX II2G + IECEx Z1 Ex ia, II2D Ex tb
• Option
BO
: ATEX II1/2G + IECEx Z0/Z1 Ex ia, II2D
• Option
BQ
: ATEX II1/2G + IECEx Z0/Z1 Ex ia
• Option
BU
: ATEX II2G + IECEx Z1 Ex ia
• Option
C2
: CSA C/US IS Cl. I, II, III Div. 1
• Option
85
: ATEX II2G + IECEx Z1 Ex ia + CSA C/US
IS Cl. I, II, III Div. 1
U
i
= 16.24 V
I
i
= 623 mA
P
i
= 2.45 W
L
i
= 0 µH
C
i
= 6 nF
For an overview and for information on the interdependencies between the gas group - sensor - nominal
diameter, see the "Safety Instructions" (XA) document for the measuring device
Low flow cut off
The switch points for low flow cut off are user-selectable.
Galvanic isolation
The following connections are galvanically isolated from each other:
• Outputs
• Power supply
Protocol-specific data
HART
Manufacturer ID
0x11
Device type ID
0x4A
HART protocol revision
7.0
Device description files
(DTM, DD
)
Information and files under:
HART load
Min. 250 Ω