Proline Promass I 100
12
Hauser
Operating tool
• Via digital communication:
– HART protocol
– PROFIBUS DP
– Modbus RS485
– EtherNet/IP
• Via service interface
Plain text display
With information on cause and remedial measures
Additional information on remote operation (→ 74)
Web browser
Plain text display
With information on cause and remedial measures
Light emitting diodes (LED)
Status information
Status indicated by various light emitting diodes
The following information is displayed depending on the device version:
• Supply voltage active
• Data transmission active
• Device alarm/error has occurred
• EtherNet/IP network available
• EtherNet/IP connection established
Ex connection data
These values only apply for the following device version:
Order code for "Output", option
M
: Modbus RS485, for use in intrinsically safe areas
Safety Barrier Promass 100
Safety-related values
Terminal numbers
Supply voltage
Signal transmission
2 (L-)
1 (L+)
26 (A)
27 (B)
U
nom
= DC 24 V
U
max
= AC 260 V
U
nom
= DC 5 V
U
max
= AC 260 V
Intrinsically safe values
Terminal numbers
Supply voltage
Signal transmission
20 (L-)
10 (L+)
62 (A)
72 (B)
U
o
= 16.24 V
I
o
= 623 mA
P
o
= 2.45 W
For IIC
*
: L
o
= 92.8 µH, C
o
= 0.433 F, L
o
/R
o
= 14.6 H/Ω
For IIB
*
: L
o
= 372 µH, C
o
= 2.57 F, L
o
/R
o
= 58.3 H/Ω
*
The gas group depends on the sensor and nominal diameter.
For an overview and for information on the interdependencies between the gas group - sensor - nominal
diameter, see the "Safety Instructions" (XA) document for the measuring device