Proline Promass 83 FOUNDATION Fieldbus
10 Technical data
Hauser
97
Zero point stability (Promass H, I):
F06-83xxxxxx-05-xx-xx-xx-000
Fig. 38:
Maximum measured error in % of reading (example: Promass 83 F / DN 25)
Calculation example (mass flow, liquid):
Given: Promass 83 F / DN 25, flow = 8000 kg/h
Max. measured error: ±0.10% ± [(zero point stability / measured value) x 100]% o.r.
Max. measured error
→
±0.10% ±
= ±0.107%
Density (liquid)
• After field density calibration or under reference conditions:
Promass F: ±0.0005 g/cc
Promass M, E, A, H: ±0.0010 g/cc
Promass I: ±0.0020 g/cc
• Special density calibration (optional), not for high-temperature version
Calibration range = 0.8...1.8 g/cc, 5...80 °C:
Promass F: ±0.001 g/cc
Promass M, A, H: ±0.002 g/cc
Promass I: ±0.004 g/cc
• Standard calibration:
Promass F: ±0.01 g/cc
Promass M, E, A, H, I: ±0.02 g/cc
Temperature
±0.5 °C ±0.005 x T (T = fluid temperature in °C)
DN
Maximum full scale value
[kg/h] or [l/h]
Zero point stability
Promass H
[kg/h] or [l/h]
Zero point stability
Promass I
[kg/h] or [l/h]
8
2000
0.20
0.20
15
6500
0.65
0.65
15 *
18000
–
1.8
25
18000
1.8
1.8
25 *
45000
–
4.5
40
45000
4.5
4.5
40 *
70000
–
7.0
50
70000
7.0
7.0
* DN 15, 25, 40 “FB” = Full bore versions of Promass I
0
±0.5
±0.2
±1.0
[%]
0
1
2
4
6
8
10
12
14
16
18
t/h
0.54 kg/h
8000 kg/h
--------------------------
100 %
⋅
Summary of Contents for Proline Promass 83
Page 24: ...3 Installation Proline Promass 83 FOUNDATION Fieldbus 24 Endress Hauser ...
Page 34: ...4 Wiring Proline Promass 83 FOUNDATION Fieldbus 34 Endress Hauser ...
Page 64: ...6 Commissioning Proline Promass 83 FOUNDATION Fieldbus 64 Endress Hauser ...
Page 66: ...7 Maintenance Proline Promass 83 FOUNDATION Fieldbus 66 Endress Hauser ...
Page 90: ...9 Trouble shooting Proline Promass 83 FOUNDATION Fieldbus 90 Endress Hauser ...
Page 116: ...Index Proline Promass 83 FOUNDATION Fieldbus 116 Endress Hauser ...
Page 118: ...www endress com worldwide BA065D 06 en 12 04 50098623 FM SGML 6 0 ...