10 Technical data
Proline Promass 83 FOUNDATION Fieldbus
92
Hauser
Measuring ranges for liquids (Promass H, I):
Measuring ranges for liquids (Promass A):
Measuring ranges for liquids (Promass E):
Measuring ranges for gases (except Promass H)
The full scale values depend on the density of the gas. Use the formula below to calculate the full
scale values:
g
max(G)
=
g
max(F)
⋅
ρ
(G)
/ x [kg/m
3
]
g
max(G)
= Max. full scale value for gas [kg/h]
g
max(F)
= Max. full scale value for liquid [kg/h]
ρ
(G)
= Gas density in [kg/m
3
] for process conditions
x = 160 (Promass F DN 8...100, M, I); x = 250 (Promass F DN 150...250); x = 225 (Promass E);
x = 32 (Promass A)
Here,
g
max(G)
can never be greater than
g
max(F)
DN
Range for full scale values (liquids)
g
min(F)
...
g
max(F)
8
0...2000 kg/h
15
0...6500 kg/h
15 *
0...18000 kg/h
25
0...18000 kg/h
25 *
0...45000 kg/h
40
0...45000 kg/h
40 *
0...70000 kg/h
50
0...70000 kg/h
* DN 15, 25, 40 “FB” = Full bore versions of Promass I
DN
Range for full scale values (liquids)
g
min(F)
...
g
max(F)
1
0...20 kg/h
2
0...100 kg/h
4
0...450 kg/h
DN
Range for full scale values (liquids)
g
min(F)
...
g
max(F)
8
0...2000 kg/h
15
0...6500 kg/h
25
0...18000 kg/h
40
0...45000 kg/h
50
0...70000 kg/h
Summary of Contents for Proline Promass 83
Page 24: ...3 Installation Proline Promass 83 FOUNDATION Fieldbus 24 Endress Hauser ...
Page 34: ...4 Wiring Proline Promass 83 FOUNDATION Fieldbus 34 Endress Hauser ...
Page 64: ...6 Commissioning Proline Promass 83 FOUNDATION Fieldbus 64 Endress Hauser ...
Page 66: ...7 Maintenance Proline Promass 83 FOUNDATION Fieldbus 66 Endress Hauser ...
Page 90: ...9 Trouble shooting Proline Promass 83 FOUNDATION Fieldbus 90 Endress Hauser ...
Page 116: ...Index Proline Promass 83 FOUNDATION Fieldbus 116 Endress Hauser ...
Page 118: ...www endress com worldwide BA065D 06 en 12 04 50098623 FM SGML 6 0 ...