9 Trouble-shooting
Proline Promag 51
86
Hauser
032
S: SENSOR SW DAT
$
: # 032
Sensor:
Error accessing the calibration
values stored in the S-DAT.
1.
Check whether the S-DAT is cor-
rectly plugged into the amplifier
board
2.
Replace the S-DAT if it is defec-
tive. Spare parts
Before replacing the DAT, check
that the new, replacement DAT
is compatible with the measuring
electronics. Check the:
– Spare part set number
– Hardware revision code
3.
Replace measuring electronics
boards if necessary.
Spare parts
Custody transfer mode:
To eliminate fault, lead seal must be
broken!
No. # 1xx
→
Software error
101
S:
GAIN ERROR AMP.
$
: # 101
Gain deviation compared to refer-
ence gain is greater than 2%.
Replace the amplifier board. Spare
parts
Custody transfer mode:
To eliminate fault, lead seal must be
broken!
111
S: CHECKSUM TOTAL
$
: # 111
Totalizer checksum error
1.
Restart the measuring device
2.
Replace the amplifier board if
necessary. Spare parts
Custody transfer mode:
To eliminate fault (point 2), lead seal
must be broken!
121
S: A / C COMPATIB.
$
: # 121
Due to different software versions,
I/O board and amplifier board are
only partially compatible (possibly
restricted functionality).
!
Note!
– The indication on the display as
notice message appears only for
30 seconds (with listing in “Previ-
ous system condition” function).
– This condition can occur if only
one electronics board has been
exchanged; the extended soft-
ware functionality is not availa-
ble. The previously existing soft-
ware functionality is still working
and the measurement possible.
Module with lower software version
has either to be actualizied by ToF
Tool - Fieldtool Package with the
required software version or the
module has to be replaced.
Spare parts
Custody transfer mode:
To eliminate fault, lead seal must be
broken!
No. # 2xx
→
Error in DAT / no communication
251
S: COMMUNICATION I/O
$
: # 251
Internal communication error on the
amplifier board.
Remove the amplifier board.
Spare parts
Custody transfer mode:
To eliminate fault, lead seal must be
broken!
No.
Error message / Type
Cause
Remedy / spare part
Summary of Contents for Proline Promag 51
Page 4: ...QUICK SETUP commissioning Proline Promag 51 4 Endress Hauser...
Page 10: ...1 Safety instructions Proline Promag 51 10 Endress Hauser...
Page 52: ...4 Wiring Proline Promag 51 52 Endress Hauser...
Page 80: ...7 Maintenance Proline Promag 51 80 Endress Hauser...
Page 104: ...9 Trouble shooting Proline Promag 51 104 Endress Hauser...
Page 122: ...11 Index Proline Promag 51 122 Endress Hauser...
Page 124: ...www endress com worldwide BA080D 06 en 03 05 50101924 FM SGML6 0...