Commissioning
Proline 800 Cellular radio
42
Hauser
8.4.2
Write protection via write protection switch
1
Off
On
2
1
Display
W
rite
protection
Custody transfer
A0047361
‣
Information regarding the
write protection switch is
provided on the connection
nameplate in the connection
compartment cover.
Unlike parameter write protection via a user-specific access code, this allows write access to
the entire operating menu to be locked.
The parameter values are now read only and cannot be edited any more.
The following parameters can always be modified even if parameter write protection is
activated:
• Enter access code
• Contrast display
• Clientt ID
1. Set the write protection (WP) switch on the display module to the
ON
position.
Hardware write protection is enabled.
In the
Locking status
parameter, the
Hardware locked
option is displayed.
On the local display, the symbol appears in the header.
V
.
8.3
m /min
3
A0044218
2.