Proline 800 Cellular radio
Electrical connection
Hauser
23
3.
5
7
4
37
E1
E2
GND
E
42 41
A0032103
3
Sensor: connection module
1. Loosen the securing clamp of the housing cover.
2. Unscrew and lift off the housing cover.
3.
NOTICE
For conduit extensions:
‣
Fit O-ring on cable and push it back sufficiently. When inserting the cable, the O-ring
must be located outside the conduit extension.
Push the cable through the cable entry. To ensure tight sealing, do not remove the
sealing ring from the cable entry.
4. Strip the cable and cable ends. In the case of stranded cables, also fit ferrules
(Verweisziel existiert nicht, aber @y.link.required='true')
.
5. Connect the cable according to the terminal assignment .
6. Firmly tighten the cable glands.
7.
L
WARNING
Housing degree of protection may be voided due to insufficient sealing of the
housing.
‣
Screw in the screw without using any lubricant. The threads on the cover are coated
with a dry lubricant.
Reverse the procedure to reassemble the sensor.