Proline 800 Cellular radio
Electrical connection
Hauser
33
5.8.1
Connection example, standard scenario
Metal, grounded pipe
A0044266
6
Potential equalization via measuring tube
5.8.2
Connection example in special situations
Unlined and ungrounded metal pipe
This connection method also applies in situations where:
• The customary potential equalization is not used
• Equalizing currents are present
Ground cable
Copper wire, at least 6 mm
2
(0.0093 in
2
)
DN
300
≤
DN
350
≥
A0029338
7
Potential equalization via ground terminal and pipe flanges
1. Connect both sensor flanges to the pipe flange via a ground cable and ground them.
2. If DN ≤ 300 (12"): Mount the ground cable directly on the conductive flange coating of
the sensor with the flange screws.
3. If DN ≥ 350 (14"): Mount the ground cable directly on the metal transport bracket.
Observe screw tightening torques: see the Sensor Brief Operating Instructions.