MMP20 SONO-DIS (option D)
Hauser
Page 51
Then set the percentage moisture of the upper moisture
value using the "Up"
and "Down"
keys, and again
press the "Measure" key
to confirm your entries.
To start measuring the material, press the "Measure"
key
again. Four measurements are taken to increase
the accuracy. An average of these measurement values is
then calculated. The measuring time is approximately
20 seconds. When the measurement is finished, the
measured pulse transit time is displayed briefly.
NOTE
: Before you start measuring, make sure that the rods
of the probe are fully immersed in the material to
be measured. The probe must remain in the
material for the entire duration of the
measurement and should not be moved.
CAUTION:
If you select "SAVE" at the end of the calibration, you overwrite one of the
pre-configured calibrations (or one you may have already modified) in the probe!
The original calibrations can only be restored with a PC and the Pico-Config software,
as well as an RS485 adapter (e.g. SM-USB).
After this, the calibration can be saved to the calibration
memory set at the start ("Save"). The selected memory is
only overwritten once the "Measure" key
is pressed.
The word "OWN:" now appears in front of the original
memory name to clearly indicate which memory has been
overwritten.
Tip: Back-up the calibrations you have saved on the probe to your PC using the "Pico-Config"
software and an RS485 adapter (e.g. SM-USB). This allows you to transfer calibrations
that you created to other probes.
14.4.3
Find probe
This menu item can be selected if there were problems communicating with the probe when the
SONO-DIS handheld device was switched on, or if a probe had not yet been connected when the
device was switched on, or if the probe is to be changed during operation. Once this item has been
selected, the SONO-DIS makes another attempt to establish a connection to the connected probe.
The serial number of the probe appears on the display as soon as the connection has been
established successfully.
"Probe not found" appears on the display if a connection cannot be established.
Note
:
If it is not possible to connect to the probe even after several attempts, check whether
the probe is connected correctly. If this does not rectify the problem, please contact our
Service Department.