MMP20 SONO-DIS (option D)
Page 48
Hauser
Menu item: "SELECT":
The 15 calibrations, and the names of the calibrations, appear on the display. This may take a
moment. Then a screen similar to that shown below appears:
Press the "Up"
and "Down"
keys to scroll through the
list and select the desired calibration. The "!" symbol in
front of a calibration indicates the currently active
calibration. Press the "Measure" key
to set the selected
calibration as the active calibration. After a brief moment,
the symbol appears in the top right-hand corner of the
display, indicating that the option has been activated.
In addition, the "!" symbol appears in front of the active c
alibration.
TIP:
You can also access this menu item directly from the measurement window by pressing the
"Up" key
.
Menu item: "CUSTOMIZE":
Here, you have the option of performing your own material calibrations or of customizing existing
calibrations to suit your particular needs. Two options are available for this:
1-point calibration:
•
Adjusts the calibration curve to the selected point.
•
The slope is not changed.
•
Just one material sample is required.
2-point calibration
•
Creation of a linear calibration between two measured
points
•
Two material samples with different moisture values are
required
1-point calibration:
This material calibration option merely adjusts (offsets) the configured calibration. As the slope
is not changed, it is important at the start to select a calibration curve that suits the material.
NOTE:
You require a sample of the material being measured to perform a 1-point material
calibration. The moisture value must be determined with another method (weighing
and drying, or similar) before the calibration.