MMP20 SONO-DIS (option D)
Hauser
Page 49
Procedure:
At the start of the calibration, the calibration memory
(01
–
15) to be overwritten must be selected with the
"Up"
and "Down"
keys. Press the "Measure" key
to accept the setting.
Then set the percentage moisture using the "Up"
and
"Down"
keys, and again press the "Measure" key
to
confirm your entries.
To start measuring the material, press the
"Measure" key
again. Four measurements are taken to
increase the accuracy. An average of these measurement
values is then calculated. The measuring time is
approximately 20 seconds. When the measurement is
finished, the measured pulse transit time is displayed
briefly.
NOTE
: Before you start measuring, make sure that the rods
of the probe are fully immersed in the material
to be measured. The probe must remain in the
material for the entire duration of the
measurement and should not be moved.
CAUTION:
If you select "SAVE" at the end of the calibration, you overwrite one of the
pre-configured calibrations (or one you may have already modified) in the probe!
The original calibrations can only be restored with a PC and the Pico-Config software,
as well as an RS485 adapter (e.g. SM-USB).
After this, the calibration can be saved to the calibration
memory set at the start ("Save"). The selected memory is
only overwritten once the "Measure" key
is pressed.
The word "OWN:" now appears in front of the original
memory name to clearly indicate which memory has been
overwritten.