Memosens COS81E
Diagnostics and troubleshooting
Hauser
25
8
Diagnostics and troubleshooting
8.1
General troubleshooting
‣
If one of the following problems is present:
Check the measuring system in the order shown.
Problem
Testing
Remedial action
Nothing displayed, no
reaction from the sensor
Power supplied to the
transmitter?
‣
Establish the power supply.
Sensor cable connected correctly?
‣
Establish correct connection.
Buildup on the spot cap?
‣
Clean the spot cap or luminescence layer
carefully with a soft cloth.
Displayed value too high
Is sensor calibrated/adjusted?
Measured value in air not 100
± 2 %SAT?
‣
Recalibrate/readjust.
When calibrating, enter the current
air pressure at the transmitter.
Displayed temperature clearly too
low?
‣
Check sensor, if necessary send sensor in
for repair.
Has salinity been taken into
account?
‣
Enter salinity value on transmitter.
Displayed value too low
Is sensor calibrated/adjusted?
Measured value in air not 100
± 2 %SAT?
‣
Recalibrate/readjust.
When calibrating, enter the current
air pressure at the transmitter.
Displayed temperature clearly too
high?
‣
Check sensor, if necessary send sensor in
for repair.
Buildup on the luminescence
layer?
‣
Clean the sensor carefully with a soft
cloth.
Fluctuations in measured
value
Are there air bubbles on the spot
cap?
1.
Change the installation angle.
2.
If necessary, change the cap type from a
u-shaped to a c-shaped cap.
Display in Vol% or %SAT
not plausible
Medium pressure has not been
taken into account
‣
Enter medium pressure on transmitter.
1. Pay attention to the troubleshooting information in the Operating Instructions for
the transmitter.
2. Check the transmitter if necessary.
Summary of Contents for Memosens COS81E