Installation
Memosens COS81E
18
Hauser
Installation position for COS81E-****U*** (with u-shaped spot cap)
A
A*
5
3
4
1
2
3
4
2
A0042966
10
Suitable and unsuitable installation positions for Memosens COS81E with u-shaped spot cap and
retractable assembly
1
Ascending pipe, best position
2
Horizontal pipe, sensor top down, impermissible due to air cushion or foam bubble forming
3
Horizontal pipe, lateral installation, with suitable installation angle
4
Upside-down installation, unsuitable
5
Down pipe, impermissible
A
Detail A (top view)
A* Detail A, turned by 90° (side view)
Possible installation angle
Inadmissible installation angle
Installation position for COS81E-****C*** (with c-shaped spot cap)
A
A*
5
3
4
1
2
3
4
2
A0042965
11
Suitable and unsuitable installation positions for Memosens COS81E with c-shaped spot cap and
retractable assembly
1
Ascending pipe, best position
2
Horizontal pipe, sensor top down, impermissible due to air cushion or foam bubble forming
3
Horizontal pipe, lateral installation with permissible installation angle (acc. to sensor version)
4
Upside-down installation, only in conjunction with c-shaped spot cap
5
Down pipe, impermissible
Possible installation angle
Inadmissible installation angle
NOTICE
Sensor not in the medium all the way, buildup, upside-down installation
These can all cause incorrect measurements!
‣
Do not install assembly at points where air pockets or bubbles may form.
‣
Avoid or regularly remove buildup on the spot cap.
‣
Do not install the sensor COS81E-****U (u-shaped) upside down.