background image

Commissioning

Memosens COL37E

10

Hauser

Refer to Operating Instructions BA02002C on the product page (

www.endress.com/cml18

)

for the measurement, calibration and adjustment of the sensor on the Liquiline Mobile

CML18.

Recommended procedure after replacing a spot cap

First calibrate and adjust the sensor at the zero point and then in the presence of oxygen.

5.1.1  

Zero point calibration

The zero point is not so important when working with relatively high concentrations of

oxygen. In these types of application, a zero point calibration is required only after the spot

cap has been replaced.
However, once oxygen sensors are used at low concentrations and in the trace range, they

must also be calibrated at the zero point.
Zero point calibrations are demanding as the ambient medium - usually air - already has a

high oxygen content. This oxygen must be excluded for zero point calibration of the sensor.
A calibration with the COY8 zero-point gel can be used for this purpose:

The COY8 oxygen-depleting gel creates an oxygen-free medium for zero point calibration.

Prior to sensor zero point calibration, check the following:

• Is the sensor signal stable?

• Has the adjustment time of 30 min - 40 min for the COY8 zero-point gel elapsed?

• Is the value displayed plausible?

1. If the sensor signal is stable:

Calibrate the zero point.

2. If necessary:

Adjust the sensor by accepting the calibration data.

If the oxygen sensor is calibrated too early, this can result in an incorrect zero point.
Rule of thumb: operate the sensor for at least 30 min in the zero-point gel.
Follow the instructions in the kit documentation enclosed with the COY8 zero-point gel.

5.1.2  

Calibration in air with 100% rH

1. Remove the sensor from the medium.
2. Clean the outside of the sensor carefully with a damp cloth.
3. Suspend the sensor just above the surface of the water. Use the calibration bottle

provided for this purpose.
Do not immerse the sensor.

4. Allow a temperature compensation time of approx. 20 minutes for the sensor in the

ambient air. Make sure that the sensor is not exposed to any direct ambient effects

(direct sunlight, drafts) during this time.

Summary of Contents for Memosens COL37E

Page 1: ...Services Operating Instructions Memosens COL37E Agile optical oxygen sensor for laboratory measurements and random sampling in the field Digital with Memosens 2 0 technology BA02236C 07 EN 01 21 00 7...

Page 2: ......

Page 3: ...ng acceptance 7 3 2 Product identification 7 3 3 Scope of delivery 8 4 Electrical connection 8 4 1 Connection to handheld device 8 4 2 Connection to handheld device via M12 cable 9 5 Commissioning 9 5...

Page 4: ...situation Failure to avoid the dangerous situation can result in a fatal or serious injury LCAUTION Causes consequences If necessary Consequences of non compliance if applicable Corrective action This...

Page 5: ...enance of the measuring system may be carried out only by specially trained technical personnel The technical personnel must be authorized by the plant operator to carry out the specified activities T...

Page 6: ...place safety As the user you are responsible for complying with the following safety conditions Installation specifications Local standards and regulations 2 4 Operational safety Before commissioning...

Page 7: ...y that it is protected against impact and moisture The original packaging offers the best protection Make sure to comply with the permitted ambient conditions If you have any questions please contact...

Page 8: ...in this window as well as the product documentation 3 2 3 Manufacturer address Endress Hauser Conducta GmbH Co KG Dieselstra e 24 D 70839 Gerlingen 3 3 Scope of delivery The scope of delivery compris...

Page 9: ...681 Connect the M12 cable to the handheld device 2 A0041680 Insert the sensor into the Memosens connection of the M12 cable and lock into place 5 Commissioning 5 1 Calibration and adjustment The senso...

Page 10: ...for this purpose The COY8 oxygen depleting gel creates an oxygen free medium for zero point calibration Prior to sensor zero point calibration check the following Is the sensor signal stable Has the...

Page 11: ...nance Take all the necessary precautions in time to ensure operational safety 6 1 Maintenance tasks 6 1 1 Cleaning of sensor Dirt on the sensor can impact the measurement and even cause a malfunction...

Page 12: ...ons for returning devices The product must be returned if repairs or a factory calibration are required or if the wrong product was ordered or delivered To ensure safe professional and swift product r...

Page 13: ...2 Zero point gel COY8 Zero point gel for oxygen and disinfection sensors Oxygen free and chlorine free gel for the verification zero point calibration and adjustment of oxygen and disinfection measuri...

Page 14: ...Oxygen gaseous hPa or Vol Temperature C F 9 1 2 measuring range 0 to 200 SAT Measuring ranges apply for 25 C 77 F and 1013 hPa 15 psi The sensor has a measuring range of up to max 1000 hPa The measure...

Page 15: ...tance NOTICE Halogen containing solvents ketones and toluene Halogen containing solvents dichloromethane chloroform ketones e g acetone pentanone and toluene have a cross sensitive effect and result i...

Page 16: ...Technical data Memosens COL37E 16 Endress Hauser Spot cap Stainless steel 1 4435 AISI 316L Spot layer Silicone 9 5 3 Temperature sensor Pt1000 Class A according to DIN IEC 60751...

Page 17: ......

Page 18: ......

Page 19: ......

Page 20: ...www addresses endress com 71558605 71558605...

Reviews: