Basic safety instructions
Memosens COL37E
6
Hauser
NOTICE
Halogen-containing solvents, ketones and toluene
Halogen-containing solvents (dichloromethane, chloroform), ketones (e.g. acetone,
pentanone) and toluene have a cross-sensitive effect and result in decreased measured values
or, at worst, in the complete failure of the sensor!
‣
Use the sensor only in media that are free from halogens, ketones and toluene.
Use of the device for any purpose other than that described, poses a threat to the safety of
people and of the entire measuring system and is therefore not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
2.3
Workplace safety
As the user, you are responsible for complying with the following safety conditions:
• Installation specifications
• Local standards and regulations
2.4
Operational safety
Before commissioning the entire measuring point:
1. Verify that all connections are correct.
2. Ensure that electrical cables and hose connections are undamaged.
3. Do not operate damaged products, and protect them against unintentional operation.
4. Label damaged products as defective.
During operation:
‣
If faults cannot be rectified:
products must be taken out of service and protected against unintentional operation.
2.5
Product safety
2.5.1
State-of-the-art technology
The product is designed to meet state-of-the-art safety requirements, has been tested, and
left the factory in a condition in which it is safe to operate. The relevant regulations and
international standards have been observed.