5. Is the measured value display on the transmitter stable:

Perform the calibration in accordance with the Operating Instructions for the

transmitter. Pay particular attention to the software settings for the stability criteria for

calibration and for the ambient pressure.

The constants K

sv

 and Tau0 of the Stern-Volmer equation are determined at both

calibration points (point in oxygen and zero point). The calibration quality index provides

an indication of the quality of the calibration in relation to the first reference calibration

of the spot cap. Therefore it is important to run the 

Change sensor cap

 command in the

calibration menu of the transmitter before every initial calibration of a spot cap.

6  

Maintenance

Take all the necessary precautions in time to ensure operational safety.

6.1  

Maintenance tasks

6.1.1  

Cleaning of sensor

Dirt on the sensor can impact the measurement and even cause a malfunction. Examples

include buildup on the spot cap, which can cause a longer response time.
The sensor must be cleaned at regular intervals for reliable measurement results. The

frequency and intensity of the cleaning process depend on the medium.
Clean the sensor:

• Before every calibration

• At regular intervals during operation as necessary

• Before returning it for repair

Type of contamination

Cleaning

Salt deposits

1.

Immerse the sensor in drinking water.

2.

Then rinse it with copious amounts of water.

Dirt particles on the sensor shaft and shaft sleeve
(

not spot cap!

)

Clean sensor shaft and sleeve with water and a suitable sponge.

Dirt particles on the spot cap

Clean the spot cap with water. No mechanical cleaning.

After cleaning:
Rinse with copious amounts of clean water.

Summary of Contents for Memosens COL37E

Page 1: ...Services Operating Instructions Memosens COL37E Agile optical oxygen sensor for laboratory measurements and random sampling in the field Digital with Memosens 2 0 technology BA02236C 07 EN 01 21 00 7...

Page 2: ......

Page 3: ...ng acceptance 7 3 2 Product identification 7 3 3 Scope of delivery 8 4 Electrical connection 8 4 1 Connection to handheld device 8 4 2 Connection to handheld device via M12 cable 9 5 Commissioning 9 5...

Page 4: ...situation Failure to avoid the dangerous situation can result in a fatal or serious injury LCAUTION Causes consequences If necessary Consequences of non compliance if applicable Corrective action This...

Page 5: ...enance of the measuring system may be carried out only by specially trained technical personnel The technical personnel must be authorized by the plant operator to carry out the specified activities T...

Page 6: ...place safety As the user you are responsible for complying with the following safety conditions Installation specifications Local standards and regulations 2 4 Operational safety Before commissioning...

Page 7: ...y that it is protected against impact and moisture The original packaging offers the best protection Make sure to comply with the permitted ambient conditions If you have any questions please contact...

Page 8: ...in this window as well as the product documentation 3 2 3 Manufacturer address Endress Hauser Conducta GmbH Co KG Dieselstra e 24 D 70839 Gerlingen 3 3 Scope of delivery The scope of delivery compris...

Page 9: ...681 Connect the M12 cable to the handheld device 2 A0041680 Insert the sensor into the Memosens connection of the M12 cable and lock into place 5 Commissioning 5 1 Calibration and adjustment The senso...

Page 10: ...for this purpose The COY8 oxygen depleting gel creates an oxygen free medium for zero point calibration Prior to sensor zero point calibration check the following Is the sensor signal stable Has the...

Page 11: ...nance Take all the necessary precautions in time to ensure operational safety 6 1 Maintenance tasks 6 1 1 Cleaning of sensor Dirt on the sensor can impact the measurement and even cause a malfunction...

Page 12: ...ons for returning devices The product must be returned if repairs or a factory calibration are required or if the wrong product was ordered or delivered To ensure safe professional and swift product r...

Page 13: ...2 Zero point gel COY8 Zero point gel for oxygen and disinfection sensors Oxygen free and chlorine free gel for the verification zero point calibration and adjustment of oxygen and disinfection measuri...

Page 14: ...Oxygen gaseous hPa or Vol Temperature C F 9 1 2 measuring range 0 to 200 SAT Measuring ranges apply for 25 C 77 F and 1013 hPa 15 psi The sensor has a measuring range of up to max 1000 hPa The measure...

Page 15: ...tance NOTICE Halogen containing solvents ketones and toluene Halogen containing solvents dichloromethane chloroform ketones e g acetone pentanone and toluene have a cross sensitive effect and result i...

Page 16: ...Technical data Memosens COL37E 16 Endress Hauser Spot cap Stainless steel 1 4435 AISI 316L Spot layer Silicone 9 5 3 Temperature sensor Pt1000 Class A according to DIN IEC 60751...

Page 17: ......

Page 18: ......

Page 19: ......

Page 20: ...www addresses endress com 71558605 71558605...

Reviews: