Operation
Memosens CFS51
28
Hauser
L
CAUTION
Medium leaking
Risk of injury, damage to clothing and the system!
‣
Make sure that the inlet and outlet of the assembly are shut off.
‣
Make sure that automatic cleaning is switched off before performing the calibration.
NOTICE
Condensation and fouling cause incorrect calibration results!
‣
Clean the sensor, and particularly the sensor's optical window, thoroughly beforehand.
‣
Avoid condensation on the sensor.
For detailed information on transmitter settings, see the Operating Instructions for
the transmitter
Pay attention to the following conditions for the calibration:
• No condensation on the sensor or the solid state reference
• Stable temperature of the sensor and the solid state reference
• Compliance with ambient temperature ranges
• Clean optical sensor window
Start calibration
1. Select
Calibration
on the transmitter.
2. Select the fluorescence sensor.
3. Select
Fluorescence
.
4. Select
Solid state reference
.
5. Follow the instructions of the transmitter.
Function check in air:
‣
NOTICE
Objects and items of clothing in front of the optical window result in incorrect
measured values!
‣
Remove any objects below the sensor (at least 0.5 m (1.64 ft)).
1
A0048475
1
Free space
Hold the sensor in the free space.
Failed function check in air:
1. Clean the optical sensor window again.
2. Repeat the measurement procedure.