Coding
Field
Selection or range
Factory setting
Display
Info
P
Function group
CHECK
(S version only)
Settings for measuring electrode
and process monitoring
P1
Switch SCS alarm
for measuring
electrode on or off
off
on
The glass electrode is monitored
for glass breakage.
(Error no.: E071.)
SCS monitoring is inactive
during calibration.
P2
Switch SCS alarm
for reference
electrode on or off
off
on
The reference electrode is
monitored for soiling and
blocking.
(Error no.: E152.)
Only if A2 = sym.
P3
Enter SCS alarm
threshold
50.0 k
Ω
0.5 ... 100.0 k
Ω
The impedance of the reference
electrode increases with the
degree of soiling.
P4
Set PCS alarm
(live-check)
off
1h
2h
4h
This function is used to examine
the measuring signal. An alarm is
signalled if it does not change for
the period selected here.
Monitoring limits:
±0.02 pH / ±5 mV / ±0.25 %
(mean value over selected period
of time).
Factory settings are printed in
bold
face;
base version does not include functions in
italics.
P
CHECK
off
P1
SCS Glas
off
P2
SCS Ref
P3
k@
SCSThresh
50.0
off
P3
PCS Alarm
Instrument configuration
Liquisys M CPM 223 / 253
32