3.5
Electrode installation and measuring cable connection
Measuring cable connection
The pH and redox electrodes are connected
using special terminated and shielded
multi-core cables.
The measuring cables (see table) can be
extended with junction box VBA or VBM (see
chpt. 9). Termination instructions are supplied
with the measuring cables.
Note:
Connectors, cable ends and
terminals are to be protected against
moisture to prevent inaccurate
measurement!
Structure and termination of measuring cables
Special measuring cables required for connection of pH / redox electrodes
Electrode type
Cable
Extension
Electrode without temperature
sensor Pt 100
CPK 1
VBA / VBM box
+ CYK 71 cable
Electrode with temperature sensor
Pt 100
CPK 7
VBA / VBM box
+ CYK 71 cable
Single pH electrode
with reference electrode and
separate temperature sensor
CPK 2
VBA / VBM box
+ PMK cable
Maximum cable length
For pH / redox
max. 50 m with CYK 71 cable
CPK127.CDR
CPK 1
CPK 2
CPK 7
PM
pH or
ORP
Outer screen
P
t
10
0
P
t
1
00
Ref.
pH or ORP
Coax bk
bn
Outer screen
gn
wt
ye
PA / PM
User must terminate end of CPK 2 cable!
bn
rd/bl
PMC
pH or redox
PMC
pH or redox
Ref. or redox
Pt 100
PM
pH + Pt 100
Coax bk
gn
bn
wt
ye
Outer screen
11
12
13
11
12
13
Reference or redox
(red inner cond.)
Fig. 3.13
Construction of special
measuring cables
CPK 1, CPK 2, CPK 7
Installation
Liquisys M CPM 223 / 253
14