Liquiline System CA80TP
Operation options
Hauser
37
6.2.2
Locking or unlocking operating keys
Locking operating keys
1. Press the navigator for longer than 2 s.
A context menu for locking the operating keys is displayed.
You have the choice of locking the keys with or without password protection. "With
password" means that you can only unlock the keys again by entering the correct
password. This password is set here:
Menu
/
Setup
/
General settings
/
Extended
setup
/
Data management
/
Change lock password
.
2. Choose whether you want to lock without or without a password.
The keys are locked. No more entries can be made. In the soft key bar, you will see
the symbol.
The password is 0000 when the device is delivered from the factory.
Make sure to note
down any changes to the password
, as otherwise you will not be able to unlock the
keypad yourself.
Unlocking operating keys
1. Press the navigator for longer than 2 s.
A context menu for unlocking the operating keys is displayed.
2. Select
Key unlock
.
The keys are unlocked immediately if you did not choose to lock with a password.
Otherwise you are asked to enter your password.
3. Only if keypad is password-protected: enter the right password.
The keys are unlocked. It is possible to access the entire onsite operation again. The
symbol is no longer visible on the display.
6.3
Configuration options
6.3.1
Display only
• You can only read the values but cannot change them.
• Typical read-only values are: analyzer data, sensor data and system information
• Example:
Menu
/
Setup
/
Analyzer
/../
Measuring parameter
6.3.2
Picklists
• You receive a list of options. In a few cases, these also appear in the form of multiple choice
boxes.
• Usually you just select one option; in rare instances you select one or more options.
• Example:
Menu
/
Setup
/
General settings
/
Temperature unit