Liquiline System CA80NO
Operation
Hauser
75
Menu/Setup/Sample preparation
Function
Options
Info
Function
Selection
• On
• Off
Factory setting
On
Diag. code 337
Read only
Warning limit
Selection
01
‐
00 to 99
‐
00 (DD
‐
HH)
Factory setting
60-00
Reset settings
Resets all settings specific to sample
preparation. All other settings are retained.
If the installation type selected is
Pressurized pipe sampling
(with e.g. Liquiline System CAT810):
Sample preparation 1
In the case of two-channel devices, sample
preparation 2 is also displayed.
Cleaning valve
Selection
• Available
• Not available
Factory setting
Not available
In the event of retrofitting, the valve can be
activated here.
If the option selected for the cleaning valve is
Available
Filter backflush
Selection
• On
• Off
Factory setting
On
If the option selected for filter backflush is
On
:
Cleaning interval
Selection
00:10 to 02:00 HH:MM
Factory setting
00:30
Compressed-air or water cleaning (optional)
for extended filter maintenance intervals
Cleaning duration
Selection
10 to 30 s
Factory setting
10 s
Cleaning duration of the automatic water or
compressed air cleaning system
Cleaning discard time
Selection
0 to 1800 s
Factory setting
180 s
Discard time once cleaning is complete. If
water is used as the flushing medium, this
must be replaced by a fresh sample before
starting the next measurement.
Reset settings
Resets all settings specific to sample
preparation. All other settings are retained.