Technical data
Liquiline System CA80NO
166
Hauser
Measuring interval
Continuous (approx. 7 min), adjustable 10 min to 24 h
Sample requirement
22 ml (0.74 fl.oz)/measurement
Reagent requirement
• Approx. 100 μl per reagent and measurement
• Given a measuring interval of 10 min, one set of reagents lasts for approx. 2 months
Standard requirements
Given a calibration interval of 48 h approx. 420 ml (14.2 fl.oz) per month
Calibration interval
1 h to 90 days, depending on the application and ambient conditions
Maintenance interval
Every 3 to 6 months, depending on the application
Maintenance effort
• Weekly: visual inspection
• Quarterly: 1 hour
15.8 Environment
Ambient temperature
range
All housing versions except "Outdoor" version
+5 to +40 °C (41 to 104 °F)
"Outdoor" version
-20 to +40 °C (-4 to 104 °F)
Storage temperature
–20 to 60 °C (–4 to 140 °F)
Humidity
10 to 95 %, non-condensating
Degree of protection
IP55 (cabinet, analyzer stand), TYPE 3R (cabinet, analyzer stand)
Electromagnetic
compatibility
Interference emission and interference immunity as per EN 61326-1:2013, Class A for
Industry
Electrical safety
According to EN/IEC 61010-1:2010, Class I equipment
Low voltage: overvoltage category II
For installations up to 2000 m (6500 ft) above MSL
Degree of contamination
Pollution level 2
8)
Sufficient mains quality is required to operate the product as designated.